PRS-812 Resistance Meter
29
Register stores up to 80 data points (MEM # Displayed after RESET)
Response &
Response from >0.1 to <1.0E06 Ohms: <2.0 seconds
Electrification:
Average Measurement Period from 0.1 ohms to 1.0E12 Ohms
2.5
Seconds.
Calculated Electrification Period per ESD S11.11,
7.5
second 0.1 ohms to
10E+12 Ohms. Programmed Electrification >1.0E+12 Ohms:
15.0+
seconds
Power :
Two 9-VDC alkaline batteries. Nominal battery life 25 hours.
Dimensions:
4.0"
wide
x 6.0" long x 2.0" deep.
Weight:
22 ounces, with batteries
Open Circuit
Current (I):
<4ma @ 100 Volts
PRS-812 Instrument Controls
FUNCTION/ MODE
Toggles Through Six (6) Operation Modes:
MODE
DISPLAYED
UNITS
INDICATION
RESIST.
RANGE
TEST
VOLTS
TEST FUNCTIONS
AUTOMATIC 1
1.0EXX
IND
- T
AUTO
AUTO. AUTO.
AUTOMATIC
RESISTANCE RANGE,
TEST VOLTS,
ELECTRIFICATION,
DISPLAY
HOLD
AUTOMATIC 2
OHMS
- T
AUTO
AUTO.
AUTO.
SEE TABLE NOTE #1
MANUAL 1
1.0EXX
IND
- T
MANUAL
MAN. MAN.
or
AUTO
TEST ONLY; NO
DISPLAY
HOLD
UL
= UNDER DECADE
LEVEL
MANUAL 2
OHMS
- T
MANUAL
MAN. MAN.
or
AUTO
OL
= OVER DECADE
LEVEL
SEE TABLE NOTE #2
MANUAL/
AUTOMATIC 1
1.0EXX
IND
- T
AUTO
&
MANUAL
MAN.
Start
AUTO
Run
AUTO.
MANUAL SETUP
STARTING DECADE,
AUTO ADJUST RANGE,
TEST VOLTS,
ELECTRIFICATION
DISPLAY
HOLD
MANUAL/
AUTOMATIC 2
OHMS
- T
AUTO
&
MANUAL
MAN.
Start
AUTO
Run
AUTO.