PRF-911 Concentric Ring
8
5. Confirming the Center Electrode
Use the following procedure to confirm proper connection to the fixture's center
electrode (D
1
) through the installed BNC/Dual Banana adapter. NOTE: Review
resistance instrumentr operating instructions before proceeding:
a. Place PRF-911 Concentric Ring fixture on a flat surface with the
electrodes facing UP.
b. Plug one PRS-801 or PRS-812 Resistance Instrument test lead the RED
BNC/Dual Banana adapter receptacle.
c. Attach an alligator clip to the second PRS-801 or PRS-812 test lead. NOTE: TO
AVOID DAMAGE, DO NOT ATTACH ALLIGATOR CLIPS TO ANY
SURFACE OR PORTION OF THE PRF-911 FIXTURE.
d. Hold the second PRS-801 or PRS-812 test lead by the insulated collar just
behind the alligator clip.
e. GENTLY touch the tip of the alligator clip to the spring loaded center
electrode. (see figure 6)
f. Select Ohms Mode by pressing the “Mode” button once on the PRS-801 or PRS-
812 Resistance Instrument.
1. Push Test Button
2. Resistance Should be < 10
Ω
Figure 5: Confirming Fixture Ground & BNC Connection