Prostat PFM-711A Скачать руководство пользователя страница 6

b.  Cycle the 

SAMPLE/HOLD

 bu  on (s2) by pushing it a few   mes; note the 

HOLD

 indica  on 

on the meter’s display. Release 

SAMPLE/HOLD

 so that the meter is free to measure fi elds.

c.  Hold a fl at sheet of material in front of the meter’s sensing plate to confi rm opera  on of 

the red ranging Bull’s Eye light system. A circle with a spot in the middle should clearly ap-
pear on the material surface when held one inch from the meter’s sensing plate.

d.  Point the unit away from surfaces and objects, and rotate the Black Panel (r16) 

ZERO

 knob 

un  l the display indicates 0.00

e.  Experiment with the meter by measuring the surface of a charged material.

f.  Slide the three way 

RANGE/POWER

 Switch (s1) to the 

V/inch

 posi  on. Allow the display to 

se  le to a stable indica  on; this should require four to eight seconds.

g.  Point the unit away from surfaces and objects, and rotate the Black Panel 

ZERO

 knob (r16) 

un  l the display indicates 000.

h.  Experiment with the meter by measuring the surface of a charged material.

OPERATIONAL NOTE

The case of PFM-711A Field Meter is conduc  ve and part of the instrument’s 

grounding circuit. If one holds the meter in one hand, and a  empts to measure 

voltage on the other hand, no reading can be obtained regardless of body voltage.

C.  Confi rming PFM-711A Field Meter Analog Func  on.

1.  Connect the analog cable to the PFM-711A 

kV/inch

 output jack.

2.  Connect the RED output banana plug to the Posi  ve (+) input of an X-Y plo  er or digital Mul  -

meter (DMM).

3.  Connect the BLACK output banana plug to the Nega  ve (-) input of the X-Y plo  er or Mul  me-

ter (DMM).

When the PFM-711A displays 1,000 volts, the analog should be approximately 0.1 volts.

IV.  Applica  on

The PFM-711A Field Meter is an accurate electrosta  c fi eld measuring device. Used by itself, the PFM-
711A will measure electrosta  c fi elds emana  ng from virtually any fl at surface or object. In 

V/inch

 mode, 

the digital display covers the range of 0 ±1999 volts in one volt increments. In 

kV/inch

 mode, it covers the 

range of 10 volts to 20kV in ten volt increments at one inch distance from the charged surface.

V.  General Opera  on

 

 

 

A.  Field Measurements using the PFM-711A Field Meter

 

1.  Before opera  ng the PFM-711A, the operator should select and wear a fully func  onal wrist 

strap that is properly grounded.

6

Rev. D  / October 4, 2018

PFM-711A Electrosta  c Field Meter

Содержание PFM-711A

Страница 1: ...User Manual PFM 711A ELECTROSTATIC FIELD METER ...

Страница 2: ......

Страница 3: ...nce User Adjustment 8 General Specifica ons 10 Table of Contents Copyright 2011 by Prostat Corpora on All rights reserved Printed in the United States of America No part of this manual may be used or reproduced in any manner whatsoever without wri en permission For informa on contact Prostat Corpora on 1072 Tower Lane Bensenville IL 60106 USA Prostat is the registered trademark of Prostat Corpora o...

Страница 4: ...pper design and subsequently the instrument is useful for making electrosta c field measurements in areas where ionized air is present 3 For accurate repeatable performance the PFM 711A must be grounded during normal opera ons II Cau ons Warnings A Do not point the PFM 711A Field Meter Ranging Lights into anyone s eyes CAUTION Do not point ranging lights into anyone s eyes B Do not store or use ins...

Страница 5: ...ent 2 Before opera ng the PFM 711A the operator should select and wear a wrist strap that is confirmed func onal and properly grounded OPERATIONAL REQUIREMENT Case of PFM 711A Field Meter MUST be grounded for proper opera on 3 With the operator s fingers contac ng the grounding snap on the back of the unit slide the three posi on RANGE POWER Switch s1 kV inch OFF V inch to the kV inch posi on a Confi...

Страница 6: ... and part of the instrument s grounding circuit If one holds the meter in one hand and a empts to measure voltage on the other hand no reading can be obtained regardless of body voltage C Confirming PFM 711A Field Meter Analog Func on 1 Connect the analog cable to the PFM 711A kV inch output jack 2 Connect the RED output banana plug to the Posi ve input of an X Y plo er or digital Mul meter DMM 3 C...

Страница 7: ... measured Move the instrument toward or away from the surface un l the RED Bulls Eye is focused on the test surface CAUTION Do not shine the RED Bull s Eye Ranging lights into yours or anyone else s eyes 4 Once the instrument is properly posi oned the PFM 711A will indicate the electrosta c field measured on its display If desired push the SAMPLE HOLD bu on s2 once to lock the indica on on the disp...

Страница 8: ...ssary that the person holding the meter be properly grounded or the meter has a ground connec on made to the metal snap on the case 9 Take An Electrosta c Field Measurement with the PFM711 A Field Meter Turn the meter ON to the desired range Hold the front of the meter approximately one inch from the surface to be measured Adjust the distance from the test surface un l the RED rang ing lights form...

Страница 9: ...ns the warranty is voided 2 Adjus ng PFM 711A Field Meter Front Panel Zero to Range The front Black panel Zero knob r16 adjustment range is less than a full turn Occasionally it may need to be adjusted to its midpoint due to environmental temperature varia ons a Select the kV inch range using the three posi on Range Power opera on switch s1 b Turn the Black Panel knob r16 fully counter clockwise c...

Страница 10: ...d Bulls Eye at 1 distance Meter Display 3 digit Liquid Crystal Display with 0 4 digit height Automatic polarity HOLD and LOW BATTERY indicators Response Digital display updates three 3 times per second Analog Output Time Constant Either Range 0 1 sec Display Time Constants High Range 0 2 sec Low Range 2 0 sec Output Analog signal output and 30 two conductor cable provided 10 0 kV meter reading equ...

Страница 11: ......

Страница 12: ...ade names are the property of Prostat Corpora on All other trademarks and trade names are the property of their respec ve companies P R O F E S S I O N A L S T A T I C C O N T R O L P R O D U C T S Prostat Corpora on 630 238 8883 Fax 630 238 9717 1 855 STATIC1 www prostatcorp com ...

Отзывы: