The PFM-711A is an accurate, compact electrosta  c fi eldmeter used for loca  ng and measuring sta  c 

charge poten  als. Other types of electrosta  c measurements may also be made using the PFM-711A and 
its accessories.

A.  The PFM-711A features 

SAMPLE

 and 

HOLD

 measurement modes that allow measurements to be 

made in places that would be diffi

  cult to reach or see with other instruments.

B.  Range switching is provided with the PFM-711A. In 

V/inch

 mode, the digital display covers the 

range of 0 ±1999 volts in one volt increments. In kV/inch mode, it covers the range of 10 volts to 
20kV in ten volt increments.

 

1.  An analog output jack provides a 1 volt output corresponding to 10kV, which is independent of 

the range selected and meter indica  on.

2.  The circuitry is a digital, electronic chopper design and subsequently, the instrument is useful 

for making electrosta  c fi eld measurements in areas where ionized air is present.

3.  For accurate, repeatable performance, the PFM-711A must be grounded during normal  

 

opera  ons.

 

 

II.  Cau  ons & Warnings

A.  Do not point the PFM-711A Field Meter Ranging Lights into anyone’s eyes.

CAUTION

Do not point ranging lights into anyone’s eyes.

B.  Do not store or use instruments in damp environments. 

C.  The PFM-711A Field Meter case must be grounded for proper opera  on.

CAUTION

Ground Case of PFM-711A Field Meter for proper opera  on.

D.  Turn OFF all instruments when in storage or transit.  Remove ba  eries when instruments are in-

tended for long-term storage, i.e. six months or longer.

 

 

III.  Equipment Inspec  on & Ini  al Prepara  on for use

A.  PFM-711A Field Meter – Equipment and Accessories

 

•  One PFM-711A Field Meter
•  One 9 Volt DC ba  ery
•  One 36 inch analog output test cable equipped with two male banana plug

4

Rev. D  / October 4, 2018

PFM-711A Electrosta  c Field Meter

Содержание PFM-711A

Страница 1: ...User Manual PFM 711A ELECTROSTATIC FIELD METER ...

Страница 2: ......

Страница 3: ...nce User Adjustment 8 General Specifica ons 10 Table of Contents Copyright 2011 by Prostat Corpora on All rights reserved Printed in the United States of America No part of this manual may be used or reproduced in any manner whatsoever without wri en permission For informa on contact Prostat Corpora on 1072 Tower Lane Bensenville IL 60106 USA Prostat is the registered trademark of Prostat Corpora o...

Страница 4: ...pper design and subsequently the instrument is useful for making electrosta c field measurements in areas where ionized air is present 3 For accurate repeatable performance the PFM 711A must be grounded during normal opera ons II Cau ons Warnings A Do not point the PFM 711A Field Meter Ranging Lights into anyone s eyes CAUTION Do not point ranging lights into anyone s eyes B Do not store or use ins...

Страница 5: ...ent 2 Before opera ng the PFM 711A the operator should select and wear a wrist strap that is confirmed func onal and properly grounded OPERATIONAL REQUIREMENT Case of PFM 711A Field Meter MUST be grounded for proper opera on 3 With the operator s fingers contac ng the grounding snap on the back of the unit slide the three posi on RANGE POWER Switch s1 kV inch OFF V inch to the kV inch posi on a Confi...

Страница 6: ... and part of the instrument s grounding circuit If one holds the meter in one hand and a empts to measure voltage on the other hand no reading can be obtained regardless of body voltage C Confirming PFM 711A Field Meter Analog Func on 1 Connect the analog cable to the PFM 711A kV inch output jack 2 Connect the RED output banana plug to the Posi ve input of an X Y plo er or digital Mul meter DMM 3 C...

Страница 7: ... measured Move the instrument toward or away from the surface un l the RED Bulls Eye is focused on the test surface CAUTION Do not shine the RED Bull s Eye Ranging lights into yours or anyone else s eyes 4 Once the instrument is properly posi oned the PFM 711A will indicate the electrosta c field measured on its display If desired push the SAMPLE HOLD bu on s2 once to lock the indica on on the disp...

Страница 8: ...ssary that the person holding the meter be properly grounded or the meter has a ground connec on made to the metal snap on the case 9 Take An Electrosta c Field Measurement with the PFM711 A Field Meter Turn the meter ON to the desired range Hold the front of the meter approximately one inch from the surface to be measured Adjust the distance from the test surface un l the RED rang ing lights form...

Страница 9: ...ns the warranty is voided 2 Adjus ng PFM 711A Field Meter Front Panel Zero to Range The front Black panel Zero knob r16 adjustment range is less than a full turn Occasionally it may need to be adjusted to its midpoint due to environmental temperature varia ons a Select the kV inch range using the three posi on Range Power opera on switch s1 b Turn the Black Panel knob r16 fully counter clockwise c...

Страница 10: ...d Bulls Eye at 1 distance Meter Display 3 digit Liquid Crystal Display with 0 4 digit height Automatic polarity HOLD and LOW BATTERY indicators Response Digital display updates three 3 times per second Analog Output Time Constant Either Range 0 1 sec Display Time Constants High Range 0 2 sec Low Range 2 0 sec Output Analog signal output and 30 two conductor cable provided 10 0 kV meter reading equ...

Страница 11: ......

Страница 12: ...ade names are the property of Prostat Corpora on All other trademarks and trade names are the property of their respec ve companies P R O F E S S I O N A L S T A T I C C O N T R O L P R O D U C T S Prostat Corpora on 630 238 8883 Fax 630 238 9717 1 855 STATIC1 www prostatcorp com ...

Отзывы: