You can change measurement type via pressing up and down buttons to select LEL, VOL, and
PPM. Once selected press MENU button. Then screen will display Range options:
You can select below values for range
for
LEL
: 0-50 or 0-100
for
VOL
: 0-5 or 0-25 or 0-30
for
PPM
: 0-10 or 0-20 or 0-30 or 0-100 or 0-200 or 0-250 or 0-300 or 0-400 or 0-500 or 0-
1000 or 0-5000
Once selected press EXIT button to return Channel Settings menu.
5- To adjust alarm levels first you have to complete measurement range settings. Then move
"Alarm levels" option on channel setting screen and select via pressing MENU button. Screen
will display default alarm level settings:
6- When press MENU button a cursor will be located under Alarm1 level. You can adjust
alarm level via using UP and DOWN buttons.
7- Then press EXIT button to release the cursor.
8- Then you can move Alarm2 via using UP or DOWN buttons.
9- Press MENU button again to activate cursor and adjust level using UP and DOWN buttons.
10- Once all completed press exit button to return the previous screen.
Alarm1 level cannot be higher than Alarm2 level. Hence it would be easier to adjust Alarm2
level first before adjusting Alarm1 level.
11- You can give a name to channel instead of default “Detectr”. Move “Define Name”
option on Channel Settings screen via UP and DOWN buttons and select via pressing MENU
button. The screen will show default name and cursor:
You can change each letter via using UP and DOWN buttons. Press MENU button to record
and move to the next letter.
12- Once all completed press EXIT button to return main menu.