Resistance 

Measurement range 

Distinguish

 

ability 

Accuracy degree 

6k

 

0.001k

 

  0.8% r3 word 

60k

 

0.01k

 

 

600k

 

0.1k

 

6M

 

0.001M

 

10M

 

0.01M

 

Overload protection: 600V DC or AC (effective value)

 

Line on-off test 

Measurement range 

Distinguish

 

ability 

Functions 

     

 

1

 

If the resistance of 
circuit being 
measured is less than 
50

±30

, then the 

beeper in the meter 
may sound. 

  Overload protection: 600V DC or AC (effective value) 

4.Operation Guide 

  4.1 Readings Hold 

During the measuring process, if the readings are required to hold, slightly 

touch 

  key ,the display value will be locked, slight press 

  key again to 

cancel readings hold. 

4.2 Backlight 

1

In the process of measurement, if the ambient light is too dim, causing reading 

difficulties, press 

  key for more than 2s, open Back Light and it will 

automatically off after about 1 minute. 

2

During the process if press 

  key for more than 2s, then the Back Light will 

be turned off. 

4.3 Auto Power Off 

1

If after 10 minutes when the meter is on without any operation, it will go into 

hibernation and automatically shut down to save power. Before the meter off, 
there will be one sound of the buzzer to remind every 1 minute. 

2

Press any key after auto power off to wake the meter into working status.

 

Содержание MT-3110

Страница 1: ...MT 3110 3 5 6 Smart Digital Clamp Meter User s Manual 1st Edition 2016 Copyright by Prokit s Industries Co Ltd ...

Страница 2: ...General protection against electric shock Prevent misuse of the meter After received the meter check if it has been damaged during the delivery After been kept and delivered in shoddy conditions check and confirm if the meter is damaged or not The pen shape meter must be in good condition Before use check the pen shape meter see if any damage to the insulation if the metal wire of the cable is bar...

Страница 3: ...to clean the meter do not use abrasive cleaning agents or solvents Power supply of the meter should be turned off when not in use range switch to the OFF position If the meter is not used for a long time batteries should be removed to prevent damage to the meter 2 Descriptions 2 1 Part Name Non contact voltage detection sensing area Data hold Back Light button Rotary switch Display screen Input so...

Страница 4: ...ndicate AUTO AutoScan Mode PMAX Maximum peak current measurement Automatic shutdown indicate LOW BATTERY Readings hold status V A Volt voltage Ampere current Ω kΩ MΩ Ohm Kilohm Megohm resistance Hz kHz Hertz Kilohertz NCV Non contact voltage detection 3 Specifications The meter should specify one year as a cycle to re calibrate in the conditions of 18 28 and relative humidity less than 75 3 1 Over...

Страница 5: ...lay symbol Temperature coefficient 0 1 accuracy degree Working temperature 18 28 Storage temperature 10 50 3 2 Technical Index AC Current Measurement range Distinguish ability Accuracy degree 6A 0 001A 2 5 reading 8 word 60A 0 01A 400A 0 1A 400A 600A 0 1A 3 reading 10 word Minimum input current 0 01A AC current Maximum input current 600A AC current Frequency range 45 65Hz When in AC current measur...

Страница 6: ...V AC Maximum input voltage 600V AC effective value Frequency range 45 65Hz Frequency Clamp head frequency measurement Pass A position Measurement range Distinguish ability Accuracy degree 60 0Hz 0 1Hz 1 0 reading 5 word 1000Hz 1Hz Measurement range 40Hz 1000Hz Input signal range 0 2A AC current effective value Pass V position Measurement range Distinguish ability Accuracy degree 60 0Hz 0 1Hz 1 0 r...

Страница 7: ...ld slightly touch key the display value will be locked slight press key again to cancel readings hold 4 2 Backlight 1 In the process of measurement if the ambient light is too dim causing reading difficulties press key for more than 2s open Back Light and it will automatically off after about 1 minute 2 During the process if press key for more than 2s then the Back Light will be turned off 4 3 Aut...

Страница 8: ...ry shall be replaced 2 When the meter has not entered into measurement it will automatically enter into the status of automatic scan the meter displays as the following diagram shows 4 5 Capture PMAX Current 1 Slightly touch PMAX key the meter displays as the following diagram 2 Hold the trigger open the clamp head and clamp one cable of the measured circuit 3 When measuring the current when captu...

Страница 9: ...se to the measured cable the meter could detect if the measured cable is 90V AC voltage or not When the meter detects AC voltage the meter buzzer alarms and the backlight flashes Note 3 Even there is no alarm indication the voltage might still exist Do not judge if there exist voltage or not in the cable relying on non contact voltage detector The detection operation might be affected by various f...

Страница 10: ...he meter display current measuring DC voltage value When measured signal is 0 5V the meter will be default as resistance value displays the ohm of the measured signal 4 9 AC Voltage Measurement 1 Connect the pen shaped meter to the measured signal when the measured signal 1 0V will the main LCD of the meter display current measuring AC voltage value and the vice LCD will display the frequency valu...

Страница 11: ... AC voltage at the same time 1 Hold the trigger open the clamp head and clamp one cable of the measured circuit when the measured signal 0 01A the meter vice LCD will display measured current value 2 Connect the pen shaped meter to the measured signal when the measured signal 1 0V the meter main LCD will display current measured AC voltage value When measured signal is 1 0V the meter will be defau...

Страница 12: ...d DC voltage value When the measured DC signal 0 5V the meter will be default as resistance value displays the ohm of the measured signal 4 13 Measure AC current and resistance at the same time 1 Hold the trigger open the clamp head and clamp one cable of the measured circuit when the measured signal 0 01A the meter vice LCD will display measured current value 2 Connect the pen shaped meter to the...

Страница 13: ...e old battery 6 Install the battery cover as previous 7 Note 8 Do not violate the battery polarity 5 2 Replace Pen shaped Meter Warnings When replacing the pen shaped meter the new ones shall be of the same or in eqaul level The pen shaped meter shall be in good condition pen shaped meter level 1000V 10A If the insulation layer of the pen shaped meter is damaged such as the metal wire of the cable...

Отзывы: