16
measured resistance is
no greater than 30
, in
approximately
40Ω-60Ω,
The red indicator will be
on.
Overload protection: 600V DC/AC
4.2.7 Capacitor
Measuring range
Resolution
Accuracy
6nF
0.001nF
± (4.0% R30 digits)
60nF
0.01nF
± (4.0% R3 digits)
600nF
0.1nF
6
F
1nF
60
F
10nF
600
F
100nF
6mF
1uF
60mF
0.01mF
± (5.0% R 3 digits)
Overload protection: 600V DC/AC
4.2.8 hFE
Functions
Value
Testing conditions
NPN or PNP
0~2000
Basic current is approx.
10uA,Vce is approx.2.8V
4.2.9 DC current
Measuring range
Resolution
Accuracy
60
A
0.01
A
±(0.8% R 3 digits)
60mA
0.01mA
600mA
0.1mA
20.00A
10mA
±(1.2% R 3 digits)
Overload protection:
protective tube for mA measuring range (FF630mA/250V) ;
protective tube for 20A measuring range (10A/250V) .
Maximal input current: mA position: 600mA DC/AC (valid value);
20A position: 20A DC/AC (valid value)
When the measured current is10A~20A, must use another 20A test lead
MT-9908.This test lead is not included
When the measured current is over 5A, the duration of continuous
measurement shall not be over 10 seconds. The current measurement shall
be carried out 1 minute after the completion of previous measurement.