2-2-2. ACV
A) Insert black test lead into the hole of “COM” and red test
lead into “
“
B) Rotate function switch to “
“, press “SELECT” button to
select the AC measurement mode.
C) The default state of the meter is in automatic range status,
which shows “AUTO” symbol
D) Put the test lead in contact with the testing point. The voltage
of the point where the red test lead is contacted will be
displayed on the screen.
Caution:
1. Don’t measure voltage over 600V. ; Otherwise, the meter will
be damaged.
2. When measuring high voltage, special attention should be
given to personal safety and avoid your body getting in touch
with high voltage circuit.
Range Accuracy Resolution
4V 1mV
40V 10mV
400V
±(0.8%+6d)
100mV
600V ±(1.0%+6d)
1V
Input Impedance: >10M
Ω
Overload Protection: 600V DC or 600V AC Peak Value
Frequency Response: (50~200) Hz
Display: Average value response (RMS of sine wave).
8