Detailed Invasive Blood Pressure Operation 

•  Connecting the Cables 

You will need a monitor specific cable to connect the SimSlim simulator IBP simulation ports to 
your  monitor.  A  variety  of  these  cables  are  available  from  Pronk  Technologies  and  other 
sources.  Pronk's cables are supplied as a 6' (1.8 m) extension cable which connects to a 6" 

(15cm) adapter cable.  Most simulator cables with 6 pin mini-din connectors should work, as the 
6 pin min-din simulator IBP pin-out is fairly standard.  Wiring is as follows: + Excit = pin 1, -Excit 
= pin 4, + Sig = pin 3, -Sig = pin 6. Contact Pronk Technologies Inc. at 800/541-9802 for more 

information. 

The IBP simulation ports are labeled on the bottom of the unit. 

•  Zeroing Simulated Blood Pressures 

In all  modes  except  IBP STEPS,  clicking the mode button will force  all four  IBP channels  to 

OmmHg  for  15  seconds.  If  you  desire  to  return  to  normal  operation  prior  to  the  15  second 
timeout, a second MODE button click will 'unzero' all channels.  (Note that this 'click to unzero 
function' is not reliable in Beta software prior to B.3) 

•  Sensitivity 

The SimSlim simulator simulates 5uVN/mmHg transducers. 

•  ECG NSR MODE 

Simulates a NSR adult patient at 60 BPM with: 

•  ART pressure on P1 of 120/80 
•  PA pressure on P2 of 30/1 O 

LV Pressure on P3 of 125/15 

•  CVP Pressure on P4 of (6) mean 

PRc:>NK  T'EOHNc:>Lc:>C31ES 

Revision 

L

 0

7/18/

2017

 

www.pronktech.com 

(800) 541-9802 

Pages 

Содержание SimSlim SL-8

Страница 1: ...SimSlim TM Patient Simulator Operator s Manual PReNK TECHNOLOGIES www pronktech com 2014...

Страница 2: ...Revision L 07 18 2017 www pronktech com 800 541 9802 Page 2 Compliant products conforming to the relevant standards and directives required for CE have a CE Mark label as shown below affixed to the u...

Страница 3: ...eration 6 Detailed Invasive Blood Pressure Operation8 Detailed Cardiac Output Operation 10 Detailed YSI Temp Operation 12 Troubleshooting Tips 13 Services 13 Warranty and Repair Information 14 Specifi...

Страница 4: ...Getting Started please refer to each parameter section in this manual for more detailed operational information 1 Turning the SlmSllm Simulator on You never have to The SimSlim simulator runs all day...

Страница 5: ...that 15 seconds After 15 seconds the blood pressures will return to their prior values You may shorten the 15 second wait time to resume IBP simulation mode by pressing the MODE button once any time a...

Страница 6: ...he snap such that the metal part of the plug is against the underside of the snap Two different size grooves are provided for different size banana plugs ECG Size ECG size is 1mV Lead II in ECG NSR MO...

Страница 7: ...1 2 MODE Revison L 07 18 2017...

Страница 8: ...t 800 541 9802 for more information The IBP simulation ports are labeled on the bottom of the unit Zeroing Simulated Blood Pressures In all modes except IBP STEPS clicking the mode button will force a...

Страница 9: ...c 200mmHg on all four channels IBP STEPS MODE In this mode a selection of static blood pressures are available on all four channels The pressures available are shown on the bottom of the unit and are...

Страница 10: ...e from the monitor Looking up the Computation Constant C C Value Once you know the current injectate temp you can look up the C C value from the table on the back of the SimSlim simulator Once you hav...

Страница 11: ...to use an injectate temp simulator rather than the actual injectate temp probe Unfortunately different monitor vendors use such a wide variety of thermistors and connectors for injectate temperature t...

Страница 12: ...1 4 1 4 monitors cable and a 3 5mm to 1 4 stereo plug converter Revision L 07 18 2017...

Страница 13: ...some monitors will occasionally ECG waveform have interaction between the IBP and ECG Unplug IBP to work around the problem On E7 hardware touching the metal shell of an IBP connector and an ECG snap...

Страница 14: ...liable only for repair or replacement of the SimSlim Simulator and optional features Pronk Technologies shall not be liable for any incidental or consequential damages Order Cancellation and Refund Po...

Страница 15: ...PRc NK TEOHNc Lc C31ES Revision L 07 18 2017 www pronktech com 800 541 9802 Page 15...

Страница 16: ...PRc NK TEOHNc Lc C31ES Revision L 07 18 2017 www pronktech com 800 541 9802 Page 16...

Страница 17: ......

Страница 18: ......

Страница 19: ......

Страница 20: ......

Отзывы: