Note!
TE module time constant
𝜏𝜏
is the time during which
TE
module temperature difference
∆𝑇𝑇
grows from 0 to 0.63
∆𝑇𝑇
𝑠𝑠𝑎𝑎
(
𝑇𝑇
𝑠𝑠𝑎𝑎
is steady state
∆𝑇𝑇
value) at electric current turned on.
The measuring procedure is carried out both for two electric supply
polarities. The data collection duration and time step can be varied.
The measuring time duration and step can be varied, too.
The measuring chart window is presented on the Figure below.
Measuring window of TEC dynamics
The obtained experimental data is then fitted by the following
function:
𝑈𝑈
𝛼𝛼
(𝑡𝑡) = 𝑈𝑈
𝑠𝑠𝑎𝑎
𝛼𝛼
(1 − 𝑒𝑒
−𝑎𝑎 𝜏𝜏
�
)
(5.2.2)
The exponential regression is based on the method of least squares.
As its outcome, the procedure provides the time constant
𝜏𝜏
and the
stationary Seebeck voltage
𝑈𝑈
𝑠𝑠𝑡𝑡
𝑈𝑈
.
page 52 of 64
ver.1.00 (2017)