![PromLegion DX4085 Скачать руководство пользователя страница 42](http://html.mh-extra.com/html/promlegion/dx4085/dx4085_user-manual_1617291042.webp)
PromLegion Ltd
Z-Meter DX4085/ User Guide
3)
mounted TE module fully described in the Z-meter database.
7.4. Measurement Procedure and Notes
Measurement cycle is started simply clicking “Measure” in the
Program Main Window.
Attention!
The temperature of a TE module changes slightly
owing to hands touching. Besides, the measuring procedure
induces a slight TE module average temperature increasing.
So keep a pause of at least 3x measured time constant
before any new measurement. It is usually 30 sec on
average. This time is usually enough to stabilize the TE
module temperature
The message:
appeared shortly after clicking “Measure” means that either:
•
open circuit inside the TE module,
•
short circuit inside the TE module
.
Note the last case is hardly possible in practice according to
PROMLEGION’s experience. Therefore, open circuit is more likely
the reason. Double-check the terminals for proper contacting TEC.
Retry measuring. Occurring of the message indicates on TE module
page 42 of 64
ver.1.00 (2017)