10.6.3 Service - pressure sensor
Mode
Menu
Archive
Reset
Service Pressure sensor
Unit Stop
Water meter
Pressure sensorFactory setting
Run. empty
Sensor Value:
Measured Value:
A0914
mm
Level sensor left
Offset:
mm
mm
Set
300
60
360
Sensor Value:
Measured Value:
mm
Level sensor right
Offset:
mm
mm
Set
300
60
360
Fig. 42: Service - pressure sensor
The
[Sensor Value]
contains the measured filling height of the storage
tank. It does not represent the actual filling height. So that the correct
filling height can be displayed, you must add the dead zone in the lower
part of the storage tank (approx. 60 mm) to the
[Sensor Value]
.
To determine the dead zone of the storage tank, you must fill the storage
tank to at least 50 %. Now you can measure the filling height with a dip
stick. You must enter this value in
[mm]
in the text field
[Measured Value]
.
The offset is calculated by the pressing the button
[Set]
.
Display of the current offset (offset = dead zone).
Measured Value = Sensor Value + Offset
Sensor Value:
Measured Value:
Offset:
Operating Menu [F2]
63