Operating instructions

Conductivity measuring cells

type LF 1 FE, LFT 1 FE, LFT K1 FE, LF 1 DE,
LFT 1 DE, LFT K1 DE, LF 1

 

1

/

2

, LFT 1 

1

/

2

, LFT K1 

1

/

2

Description

The electrode cells type LF 1 FE, LFT 1 FE, LFT K1 FE, LF 1 DE, LFT 1 DE,
LFT K1 DE, LF 1 

1

/

2

, LFT 1 

1

/

2

, LFT K1 

1

/

2

 with PG 13.5 or 

1

/

2

 screwed thread as well as

with or without integrated Pt 100/1000 for temperature compensation serve the
conductive measurement of the electrolytic conductivity in aqueous media. The sensors
are designed either with fixed cable (designation ... FE) or with DIN 4P right-angle plug
(designation ... DE). The instrument is connected to a measuring device via a 2- or
4-wire instrument lead.

IMPORTANT
The measuring cell must be placed in distilled or demineralized water for
5 - 10 minutes prior to initial operation.
For correct measurement, particular care must be taken to ensure that no
air bubbles remain in the gap (i.e. between the measurement electrodes).

Servicing/cleaning

Operation of the sensor should be checked regularly and the sensor is to be calibrated
as required. The gap between the measuring electrodes must be checked regularly for
dirt deposits. Any adhering deposits can be removed with the aid of a soft water jet, by
immersion for 2 - 3 minutes in diluted (1% solution) acid or by wiping with a cloth or soft
brush (e.g. toothbrush/bottle brush).

Storage

Dry

Installation

For installation into the flow meter type DLG (1

 boring), an adapter PG 13.5/1

 (parts

number 1002190) is required!
Particular care must be taken during installation to ensure that the measuring electrodes
are always completely covered with liquid medium.

IMPORTANT
Because of their material, the measuring cells are not suitable for
measurement of surfactant-containing cleaning agents and solvent-
containing media.

Technical Data

Cell constant:

k = 1.0 cm

-1

 (

±

 5 %)

Measuring range:

approx. 0.01 ... 20 mS/cm

Medium temperature:

0...80 

°

C (unpressurized)

Max. pressure:

16 bar (at 25 

°

C)

Thread:

PG 13.5 or 

1

/

2

Dimensions:

Stem length
(up to push-and-twist connector)
120 mm; ø 12 mm

Storage temperature:

-5...50 

°

C

Electrodes:

Special graphite

Temperature sensor:

Pt 100 or Pt 1000
(integrated in cell stem)
in type LFT/LFT K

Cell stem:

Epoxy

Electrical connection
to sensor:

fixed cable 5 m, types ... FE
DIN 4P right-angle plug,
types ... DE

Electrical connection
to measuring instrument: 2- or 4-wire instrument lead

type of enclosure IP 65

Pin Assignments

white/brown: Electrodes
green/yellow: Pt 100/Pt 1000

DIN plug connection: electrodes 

 +2, Pt 100/Pt 1000 1+3

120 

±

3

Ø 12 mm

120

pk_6_085

pk_6_086

Ø 12 mm

BA_DT_079_09_06_G_GB_F_E.p65

20.09.2006, 11:35 Uhr

2

Содержание LF 1 FE

Страница 1: ...Einbau in den Durchlaufgeber Typ DLG 1 Bohrung ist ein Adapter PG 13 5 1 Teile Nr 1002190 notwendig Beim Einbau ist darauf zu achten dass die Messelektroden immer vollständig mit Flüssigkeit bedeckt sind ACHTUNG Die Messzellen sind materialbedingt nicht geeignet für die Messung in tensidhaltigen Reinigern und lösungsmittelhaltigen Medien T Nr Part No 987823 BA DT 079 09 06 G GB F E Anschriften und...

Страница 2: ...aid of a soft water jet by immersion for 2 3 minutes in diluted 1 solution acid or by wiping with a cloth or soft brush e g toothbrush bottle brush Storage Dry Installation For installation into the flow meter type DLG 1 boring an adapter PG 13 5 1 parts number 1002190 is required Particular care must be taken during installation to ensure that the measuring electrodes are always completely covere...

Страница 3: ...cide dilué à 1 pendant 2 à 3 minutes ou en frottant avec un chiffon ou une brosse molle p ex brosse à dents brosse à bouteilles Stockage A sec Montage Un adapteur PG 13 5 1 nº de pièce 1002190 est nécessaire au montage dans le capteur continu type DLG alésage 1 Lors du montage il convient de veiller à ce que les électrodes de mesure soient toujours entièrement entourées de liquide ATTENTION Les so...

Страница 4: ...nte 2 3 minutos en ácidos diluidos 1 o frotando con un trapo o un cepillo suave cepillo de dientes cepillo para botellas Almacenamiento En lugar seco Montaje Para el montaje en un transmisor continuo del tipo DLG orificio 1 es necesario un adaptador PG 13 5 1 núm de pieza 1002190 En el montaje debe tenerse cuidado de que los electrodos de medición estén cubiertos siempre totalmente con líquido ATE...

Отзывы: