Transport and Storage
NOTE
Sensor must be transported, sent and stored in the original packaging
only. Please retain packaging complete with polystyrene parts!
Storage
Storage and transport temperature
5 to 50 ºC
Humidity
max 90 % relative humidity,
non-condensing
Storage period of sensor and
electrolyte in original packaging:
1 year
NOTE
If the sensor has been stored for longer than the stated period, please
return it to ProMinent for checking or overhaul. Reliable function and
measuring accuracy can otherwise not be guaranteed.
Contents
•
1 CDE measuring sensor with diaphragm cap and clamping ring
•
1 bottle of electrolyte (100 ml)
•
1 operating instructions
•
1 srewdriver
5
Assembly
Adding the
electrolyte
IMPORTANT
•
The diaphragm at the lower end of the diaphragm cap and the
electrodes at the lower end of the electrode shaft should not be
touched, damaged or brought into contact with greasy substances!
Otherwise the sensor will not work correctly. If this is the case,
replace diaphragm cap or return for refurbishing the electrodes.
•
Protect yourself and your clothing from contact with the electrolyte
(acid!) using appropriate protective equipment. The electrolyte can
cause irritation to skin and damage to clothing. In case of contact,
rinse immediately with plenty of cold water.
NOTE
Carry out the following steps at the washbasin!
왘
Open the electrolyte bottle and screw on the nozzle.
왘
Remove the red cap completely from the nozzle and cut off the end of the
nozzle.
왘
Remove the diaphragm cap and unscrew the diaphragm cap from the
electrode shaft.
왘
Rinse the diaphragm cap and the electrode with a small amount of
electrolyte.
왘
Fill the diaphragm cap up the top with electrolyte and allow electrolyte to
overflow down the inside wall.
Transport and Storage / Assembly
BA_DT_095_08_08_4spr.p65
05.08.2008, 12:42 Uhr
22