16
Complete Operating Menu / Description
D1C2-033-Redox-GB
relays
setting ?
control timer 1
relay 1
period
1 h
t on:
1 min
control timer 2
relay 2
period 1
h
t on:
1 min
with timer
relay adjustment
relay1: timer
relay2: timer
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Relay adjustment
as per
identity code
Relay 1
Solenoid valve 1
Limit value 1*
*For “limit value”, the relays
Actuator 1
remain active, even in the
Timer 1
event of a fault.
Servomotor
only with servomotor
off
Relay 2
Solenoid valve 2
Limit value 2*
Actuator 2
Timer 2
off
Cycle
10 s
1 s
10 s
9999 s
for solenoid valve
min. time
1 s
1 s
1 s
Cycle/2
for solenoid valve
Set here the smallest
permitted operating factor
of the connected device.
Period (Cycle)
off
1 h
1 h / off
240 h
for timer
t on
1 min
1 min
1 min
60 min
for timer
NOTE
The limit value relay can be defined in such a way as to respond as an actuator, i.e. if a limit
value relay closes a circuit, it opens when a pause contact is activated and/or for a
subsequent delay period t
d
(if t
d
is set to > 0 min in “General settings”).
Cycle
Timer relay
off
on
t
t
on
IMPORTANT
The timer will be reset if there is a drop in the power supply.
BA_DM_183_10_10_EN.p65
28.10.2010, 11:44 Uhr
16