Complete Operating Menu / Description

D1C2-033-Redox-GB

relays

setting  ?

control timer 1

relay 1

period

1 h

t on: 

1 min

control timer 2

relay 2

period 1 

h

t on: 

1 min

with timer

relay adjustment

relay1: timer

relay2: timer

Possible values

Initial value

Increment

Lower value

Upper value

Remarks

Relay adjustment

as per

identity code

Relay 1

Solenoid valve 1

Limit value 1*

*For “limit value”, the relays

Actuator 1

remain active, even in the

Timer 1

event of a fault.

Servomotor

only with servomotor

off

Relay 2

Solenoid valve 2

Limit value 2*

Actuator 2

Timer 2

off

Cycle

10 s

1 s

10 s

9999 s

for solenoid valve

min. time

1 s

1 s

1 s

Cycle/2

for solenoid valve

Set here the smallest

permitted operating factor

of the connected device.

Period (Cycle)

off

1 h

1 h / off

240 h

for timer

t on

1 min

1 min

1 min

60 min

for timer

NOTE

The limit value relay can be defined in such a way as to respond as an actuator, i.e. if a limit
value relay closes a circuit, it opens when a pause contact is activated and/or for a
subsequent delay period t

d

 (if t

d

 is set to > 0 min in “General settings”).

Cycle

Timer relay

off

on

t

t

on

IMPORTANT

The timer will be reset if there is a drop in the power supply.

BA_DM_183_10_10_EN.p65

28.10.2010, 11:44 Uhr

16

Содержание DULCOMETER D1C A D Series

Страница 1: ...850 ProMinent Dosiertechnik GmbH 69123 Heidelberg Germany BA DM 183 10 10 EN Please completely read through operating instructions Do not discard The operator shall be liable for any damage caused by...

Страница 2: ...mV Correction variable 0 None Feed forward control 0 None 1 via standard signal 0 4 20 mA 2 via frequency 0 500 Hz 3 via frequency 0 10 Hz Control input 0 None 1 Pause Signal output 0 None 1 standard...

Страница 3: ...data and function of the series DULCOMETER D1C controller provide detailed safety information and are divided into clear steps IMPORTANT Please observe the parts of these operating instructions appli...

Страница 4: ...button To change over within a menu level and to change from one variable to another within a menu point START STOP button Start stop of control and metering function ENTER button To accept confirm or...

Страница 5: ...riable starts again after cessation of pause 4 4 Feed Forward Control The D1C controller can process a signal of a feed forward control Depending on the device version see identity code and the settin...

Страница 6: ...off left Symbol Control on left Metering pump 2 reducing agent Symbol Control off right Symbol Control on right Solenoid valve 1 oxidant Symbol Control off left Symbol Control on left Solenoid valve...

Страница 7: ...Operation Access code correct Parameter setting Calibration notes Permanent display 1 Permanent display 2 Calibration menu Various Access code Setting menus D1C2 007 GB The various menus are selected...

Страница 8: ...er Ti Td xp 0 s 750 mV setpoint 1 lower 650 mV 10 ctrl parameter xp 10 control settings Only with control control positive oxide regulated value negative reduc manual dosing control current regulat va...

Страница 9: ...ys After everything has been checked all error tests which refer to the measured value are restarted Restricted Operating Menu Description Error message Condition Effect Measured value high Measured v...

Страница 10: ...value 30 control current regulat value 30 D1C2 015 Redox GB Access to all setting menus can be blocked with an access code Only with control Note The controlled variable is recalculated every second...

Страница 11: ...ode 5000 1 1 9999 Language as per identity as per identity code code Operating menu restricted restricted complete D1C2 016 GB Access to all setting menus can be blocked with an access code general se...

Страница 12: ...Only with servomotor Only with pumps meas value setting pumps setting relays setting limits setting servomotor setting Number and scope of setting menus is dependent on the device Access to setting me...

Страница 13: ...V 2000 mV if measured variables 2 and 5 185 265 mV 220 mV are connected 425 505 mV 465 mV 1000 mV if measured variable 1 is connected Complete Operating Menu Description D1C2 018 Redox GB Permanent di...

Страница 14: ...ng value input changes within the control time measuring value It is assumed that the value does not change for an intact sensor If the measuring value does not change during this control time the DUL...

Страница 15: ...ump1 pump2 control ORP pump1 oxide pump2 reduc pulse minute 180 180 In the setting menu correction value for this equipment enables you to display the temperature or to maintain an mA signal proprotio...

Страница 16: ...motor off Relay 2 Solenoid valve 2 Limit value 2 Actuator 2 Timer 2 off Cycle 10 s 1 s 10 s 9999 s for solenoid valve min time 1 s 1 s 1 s Cycle 2 for solenoid valve Set here the smallest permitted op...

Страница 17: ...n time smallest permitted operating factor of the connected device The actuating variable determines the ratio ton cycle and thus the switching times see fig above The min time influences the switchin...

Страница 18: ...Actuating direction active closed active closed Reacts as make contact limit value 1 active open Reacts as break contact limit value 2 Switch on delay t on 0 s 1 s 0 s 9999 s Switch off delay t off 0...

Страница 19: ...er value Remarks Servomotor Setting Setting ok off Control direction Oxidant Oxidant Reducing Control range 100 1 10 100 in of operating range D1C2 025 Redox GB Access to all setting menus can be bloc...

Страница 20: ...026 Redox GB 10 control load regulated range Access to all setting menus can be blocked with an access code Only with control Note The controlled variable is recalculated every second Only suitable fo...

Страница 21: ...ltiplicative multiplicative Feed forward control effect additive Max add regulated only with add feed variable 100 1 500 500 forward control D1C2 027 Redox GB Access to all setting menus can be blocke...

Страница 22: ...tting menus can be blocked with an access code general setting information ident code D1CA software version DxRxxxxxxxxxx D1C A1 FW 5 00 alarm relay access c operating menu active 5000 english complet...

Страница 23: ...ows to the sensor With PID control Identity code characteristics control characteristic 2 The control variable output resulting from the pause and the expiry of the time delay td is reconciled jointly...

Страница 24: ...error mV calibration Basic load Stop No Metering continues in Check probe replace if faulty case of error with unsteady necessary recalibrate measured values if necessary Correction measured variable...

Отзывы: