14
Possible values
Initial value
Increment
Lower value
Upper value
Remarks
Standard signal input
4 mA
0 mA
lower signal limit
4 mA
Allocated
0–1 V
1 mV
-2000 mV
+2000 mV
probe voltage
Checkout time
off
1 s
1 s
9999 s
Constant measurement
signal results in message
and alarm.
Function off = 0 s
Control Time Measuring Value
IMPORTANT
This function may not be activated for applications where it can be assumed that the
measuring value does not change.
This function checks whether the measuring value from the sensor (measuring value input) changes within
the “control time measuring value”. It is assumed that the value does not change for an intact sensor.
If the measuring value does not change during this control time, the DULCOMETER
®
D1C sets the control
output to “0” and the alarm relay drops out. The LCD display shows e.g. the message “check mV-probe”.
IMPORTANT
When changing the range adjustment, the adjustments in all menus have to be checked!
Complete Operating Menu / Description
Measured Value
D1C2-020-Redox-GB
Access to all setting menus can be blocked with an access code !
Standard signal input mA
meas. value
range adjustment
4 mA
= 0 mV
20 mA = 1000 mV
meas. value
range adjustment
meas. value
checkout time
measured value
off
meas. value
setting ?
BA_DM_183_10_10_EN.p65
28.10.2010, 11:44 Uhr
14