MAINTENANCE MANUAL //

 STRETCHER PC-750 + RAIL PB-720

Revision 2021/10

2

1.1 STRETCHER

1.1.1 CONTROL PANEL:

 

Check that the membrane is well attached to the housing to prevent moisture from entering.

1.1.2 BATTERY BOX:

 

Open the box to check that the seal is maintained. There should be no signs of moisture inside the box.

 

Make sure that the 4 closing screws of the box are correctly tightened.

1.1.3 LOCK PEDAL:

 

Check the correct operation of the pedal: operate the pedal and check that the 2 front wheels are com-

pletely locked. Release the pedal and check that the wheels rotate freely.

1.1.4 FRONT AND REAR WHEELS:

 

Check the correct operation of the wheels:

 

All 4 wheels rotate smoothly when free.

 

The 2 rear wheels are locked when the locking tabs are operated respectively.

 

The 4 wheels are braked by activating the brake pedal for each wheel.

 

The 4 wheels roll correctly and do not show any deformation.

1.1.5 LOWER STRUCTURE:

 

Check the tightening torque of the screws on both sides of the 

lower structure (see Fig. 1). The tightening torque must be 19.11 

Nm.

 

Clean the guides on both sides and renew the lubrication with 

Vaseline (see Fig. 2). IMPORTANT: after greasing, always clean the 

excess.

 

Check that the guide bearings are in good condition, and con-

sider whether they need to be replaced: during the process of 

raising and lowering the stretcher, check that the bearings roll 

correctly.

1.1.6 PROFILE:

 

Check the correct functioning of the patient’s fixation systems, 

and consider whether it is necessary to replace them:

 

They open and close properly.

 

They are not worn, torn, or have any broken parts.

 

The fixing pieces on the profile are correctly fastened.

 

Check the correct operation of the handrails, and consider whe-

ther it is necessary to replace them:

 

They open and close properly

 

The bolt of the rear support (see Fig. 3) that fixes the 

handrails on the profile is well tightened

 

Check the correct operation of the extendable handles, and con-

sider whether it is necessary to replace them:

 

They open and close properly

 

They stay locked in the extended and closed position

FIGURE 1

FIGURE 3

FIGURE 2

01 

EVERY 6 MONTHS

Содержание PB-720 Series

Страница 1: ...MAINTENANCE MANUAL ELECTRIC STRETCHER SERIES PC 750 RAIL SERIES PB 720 Revision 2021 10...

Страница 2: ...e screws on both sides of the lower structure see Fig 1 The tightening torque must be 19 11 Nm Clean the guides on both sides and renew the lubrication with Vaseline see Fig 2 IMPORTANT after greasing...

Страница 3: ...ee Fig 4 Check that all the elements screwed on top of the rail are fixed central hooks front lock loading head etc Open the manhole covers clean and grease the lateral move ment guides with Vaseline...

Страница 4: ...tor when raising and lowering the stretcher and confirm that it is working well without exerting any extra effort 2 1 5 LOADING HEAD Check the return of the load terminals of the head press each of th...

Страница 5: ...nt is not correctly assembled it could cause damage during use in which case the Promeba technical service should be contacted to consider whether it needs to be replaced Check that the screws that co...

Страница 6: ...ery two years or according to wear Check the condition of the telescopic lever handles and their inner profiles and consider whether they need to be replaced These elements must be replaced every two...

Страница 7: ...r it is necessary to replace them 4 2 RAIL 4 2 1 FIXED RAIL OPERATION Check the condition of the adhesives and consider whether it is necessary to replace them 4 2 2 FIXED RAIL OPERATION Check the con...

Страница 8: ...derecho ha modificaciones sin previo aviso Promeba S L no se considera responsabe de los da os causados por la falta o la inexactitud de la informaci n aqu mencionada All rights reserved Variations ca...

Отзывы: