USER’S MANUAL. MD-200B
03/2006
Page
3
En
g
lis
h
Electrical Characteristics
Accuracy is given as ± (measurement error + display error) and applies at 23
°
C
±5
°
and less than 70% RH.
D.C. Voltage
Range Resolution
Accuracy
Over voltage
protection
400 mV
100 µV
± 0,6 % reading ± 2 digits
4 V
1 mV
40 V
10 mV
400 V
100 mV
600 V
1 V
± 0,4 % reading ± 2 digits
1100 V DC or Vac
peak
Input impedance
10 M
Ω
, < 100pF
A.C. Voltage
Accuracy
Range Resolution
50 Hz to 60 Hz
40 Hz to 1 KHz
Over voltage
protection
4 V
1 mV
40 V
10 mV
400 V
100 mV
600 V
1 V
± 0,8 %
reading ± 5
digits
± 1,3 %
reading ± 5
digits
1100 V DC or Vac
peak
Input impedance
10 M
Ω
, < 100 pF
Conversion of AC voltages
Mean measured - Rms. value displayed
D.C. Current
Range Resolution
Accuracy
Voltage
Burder
4 mA
1 µA
40 mA
10 µA
400 mA
100 µA
± 0,75 % reading ± 2 digits
800 mV max.
10 A
10 mA
± 1,5 % reading ± 4 digits
1 V max.
10 A range
30 seconds maximum above 10 A.
Input protection
By means of fuses in both inputs.
Содержание MD-200B
Страница 1: ...MD 200B MULT METRO DIGITAL DIGITAL MULTIMETER MULTIM TRE NUM RIQUE 0 MI1393...
Страница 2: ......
Страница 4: ......
Страница 5: ...SUMARIO CONTENTS SOMMAIRE Manual espa ol English manual Manuel fran ais English Fran ais...
Страница 6: ......
Страница 8: ...MANUAL DE INSTRUCCIONES MD 200B...
Страница 14: ...MANUAL DE INSTRUCCIONES MD 200B P gina 6 03 2006...
Страница 30: ...MANUAL DE INSTRUCCIONES MD 200B P gina 22 03 2006...
Страница 32: ...USER S MANUAL MD 200B...
Страница 38: ...USER S MANUAL MD 200B Page 6 03 2006...
Страница 54: ...USER S MANUAL MD 200B Page 22 03 2006...
Страница 56: ...MANUEL D UTILISATION MD 200B...
Страница 62: ...MANUEL D UTILISATION MD 200B Page 6 03 2006...
Страница 78: ...MANUEL D UTILISATION MD 200B Page 22 03 2006...