System Configuration Preva
20
Changing Pre-programmed Exposure Settings
The Preva Dental X-Ray System allows the operator to increase or decrease
image density for all presets for a receptor simultaneously or to change each of
the technique factors for a preset individually. You can also restore factory
default settings. For charts of the factory default settings, refer to Factory Default
Exposure Settings later in this manual.
Note:
If the 12 inch cone is going to be used, configure the Preva for use with
the 12 inch cone before changing pre-programmed exposure settings.
Configuring the Preva for use with the 12 inch cone will reset exposure settings
to the default settings used with the 12 inch cone.
Note:
Before changing presets, use the table below to write down the presets
you are programming.
Displaying the
Preset Options
Menu
1. From the Main System Configuration menu, shown in Figure 3, select
CHANGE PRESETS. You will see the Preset Options menu shown in Figure
5.
2. Selecting EXIT returns the display to the Main System Configuration menu
shown in Figure 3.
Figure 5
Preset Options
Menu
Changing All
Receptor
Settings
Globally
1. Select ALTER DENSITIES from the Preset Options menu. The first Image
Receptor Type illuminates. The display shows the selected Image Receptor
Type and current density.
2. Using the Image Receptor Type button, select the image receptor to adjust.
3. Use the up and down arrow buttons to specify a percentage by which
densities will be increased or decreased for the selected receptor. Densities
can be increased or decreased according to values provided on the display.
4. Press Enter to save your settings.