Solving Performance Issues Preva
15
Solving Performance Issues
Performance Issues
Light or Dark X-
Ray Images
1. Adjust the selected exposure time, kilovoltage or tube current to produce
an acceptable image. If necessary, reprogram the techniques factors, as
explained in the System Configuration section of this manual.
2. Verify the kilovoltage and tube current during an exposure using the
diagnostic mode, as explained in the System Configuration section of this
manual. Alternatively, you may employ a non-invasive meter to evaluate
kilovoltage and exposure time.
3. Inspect the condition of the remaining imaging chain components such as
the film, chemistry and processor, or the condition of the x-ray sensor and
computer.
No X-Ray
If no x-ray is produced, check the following:
1. Verify that the line cord (if one is in use) is properly connected.
2. Verify that the power switch is in the ON position.
Pre-termination
Error
Early release of the exposure switch will cause a pre-termination error to
occur. After five seconds, the system will return to the normal operating
condition. Be advised that this will result in an underexposed image.