![Process Technologies SK 7 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/process-technologies/sk-7/sk-7_operating-manual_1608706014.webp)
SKProfilemaster
The next example below shows a 3-zone profile envelope band with different upper and lower
Sensitivity settings per zone:
Please note:
Changing the timing window will automatically erase the profile zones on the channel in question,
and you will have to go to the MENU section (page 4/4, SKProfilemaster) and re-activate the
number of profile zones you wish to have for the new timing window.
The SKProfilemaster envelope profile band.
The 3-zone setting allows for an even closer
match between curve segment variation and
adopted profiles.
How to set the SKProfilemaster band:
- touch SKProfilmaster box in the single force
curve screen to access sensitivity entry mode
- separate sensitivity windows pop up for each
individual profile zone
- touch the border lines direktly to modify
the zone sizes
- set each zone’s desired sensitivity from
1 to 9 (1=coarse; 9=fine)
- confirm each zone’s selection by pressing the E key
(you are able to view immediately how your setting is
affecting the envelope width in each zone)
Sensitivity windows
SK 7
V5.2 E 10/04
SK7-13