Purge = 10 seconds
Fill = 20 seconds
Sample = 10 seconds
Delay = 2 seconds.
These times can be adjusted to optimize the sequence to the flow rate of the product.
Changing the Sequence Times
Use the Operator Interface.
Use the down arrow to move the bottom line of the display to Miscellaneous.
Press the right arrow to enter this selection.
At the password request, set the password to 2882, then press ENTER.
This will access the Engineering menu.
The menu contains many items that if changed will influence how the sensor operates
Use the down arrow to scroll through the selections until the display shows:
SAMPLER: Y/N
Press the right arrow to make the cursor flash on Y and press ENTER to activate the sampler mode.
The sensor software will normally have been pre-set on delivery to make the sensor operate in the
sampler mode, in which case the display will show:
SAMPLER: ON
Press the down arrow to access the times for the various cycles.
SAMP PURGE = 0010
Use the right arrow to move the cursor under the digits, up/down arrows to set a new time and press
ENTER to set the time.
Press the down arrow to see the next time.
55