• You can add text notes to a defined area by tapping 10: just tap on
the desired cells, confirm with
and enter your text; the selected
area will be highlighted with a white transparent overlay. The note can
be deleted anytime later by tapping on the area and pressing
.
After storage (tap 7), the data can be seen in the Corrosion-Scan View,
Distribution View, Cumulative Distribution View, Chipping Graph View,
Chipping Graph ASTM (see “4.2.4. Corrosion Display Views”).
One wheel and four wheel electrode
1
2
3
4
5
6
7
8
10
11
14
13
12
9
15
Figure 37:
Corrosion-Scan Mode with one and four wheel electrode
1
File name:
enter the file name and tap return. Saved measurements
will be stored with this file name. If several measurements are made
under the same filename, a suffix increments after each measure-
ment and follows the file name. A file name given by the user ending
already with a number will be incremented on next measurement
based on this last number).
2
Interface box connected, current time, battery status.
3
Display
of selected Settings:
• Probe Type
• Grid X Spacing
• Grid Y Spacing
• Grid Limits (if set)
Zoom
in by placing thumb and index finger together on the
screen and spreading them apart. This can be used in both the
horizontal and vertical directions when making a measurement.
Zoom
out by placing thumb and index finger apart on the
screen and pinching them together.
Pan
the image from left to right by dragging.
The potential values associated with each cell can be visual-
ized by zooming in at the needed level.
Double tap
: zoom to fit the whole scan
4
Color palette of displayed potential values
5
Settings:
switches to the settings menu
6
Start / stop measurement
7
Store measured data
8
Rotate probe (only four wheel electrode)
in the first 1 m / 40 inch
of a new line
9
Direction button:
in the first 1 m / 40 inch of a new line it switches
the direction of the scan (default direction is defined under Direction
of Next Line setting)
10 Move to next line
11 Edit notes / Edit cell values / Reset all values:
tap cell to edit
12 Probe position and horizontal distance from start