19
© 2017 Proceq SA
2
Probe connected, current time, battery status and warning triangle
for zeroing the probe when needed: after 5 minutes in orange, after
10 minutes in red..
NOTE! Tap on the triangle to perform zeroing.
3
Display
of selected Settings:
• Measuring Range
• Type of correction (None, AI, NRC, CAL)
• 1
st
and 2
nd
layer rebar size and spacing
• Cover Offset (if set)
• Probe Direction X: Undefined direction
^, v, <, >: On vertical wall, probe head towards
up, down, left, right
_, ˉ: On horizontal surface, on soffit
Zoom
in by placing thumb and index finger together on
the screen and spreading them apart. This can be used in
both the horizontal and vertical directions when making a
measurement.
Zoom
out by placing thumb and index finger apart on the
screen and pinching them together.
Pan
the image from left to right by dragging.
4
Measured values displayed:
Cover / Diameter
5
Settings:
switches to the settings menu
6
Rebar Diameter:
measuring of rebar diameter
7
Store measured data
8
Restart: all data of current measurement are deleted
9
Delete current line
10 Move to next line / Rotate probe
11 Zoom to fit the whole scan
12 Actual Cover
13 Distance to the nearest rebar
14 Nos. / Measured diameter
15 Skip distance
16 Skip button
17 Coils indicator
18 Speed bar
NOTE! It is advisable to locate the first and second layer rebars
with the Locating Mode to find the optimum line position prior
to measuring with the Multi-Line Mode (see Figure 9). “Align
Rebar Position” should only be set, if all rebars are running
parallel to the start line (Y-axis). For large areas it is advisable
not to set “Return to start on new line” and to measure the
lines alternatively start to end, back from end to start.
• Position the probe cart at the start line in an optimum position (see
Figure 9, the MC C / D at the mid line of the rebars running parallel to
the moving direction, both rectangles in 16 are of equal size).
• Start with the measurement if the cursor is at the start line. If not per-
form a reset 8.
• Move the probe cart crosswise over the rebars. Above each rebar,
when the red LED of MC C / D lights up, you may measure the rebar
diameter and on completion, it will be displayed. If the spacing be-
tween parallel or orthogonal rebars is in the range of 5 to 30 cm (2.00
to 12.00 inch), activate AI or NRC correction. At the end of the first line
a marker (dotted blue line) is set. To proceed with the next line tap 10