7.  Technical Specifications

Uniaxial (or unconfined) Compressive Strength Range

ISRM recommend the use of rebound hammers for UCS within the range 20 – 150 MPa.

ASTM recommend an upper UCS limit of 100 MPa.

Other studies have been made on rocks with UCS up to 250 MPa.

Proceq’s experience from concrete testing recommends a lower limit of about 10 MPa and above 
100 MPa caution should be exercised due to the gradient of the correlation curve.

Mechanical Data

Impact energy

(N) 2.207 Nm, (L) 0.735 Nm 

Hammer mass

115 g 

Spring extension

75 mm (2.95”)

Plunger radius

25 mm (0.98”)

Dimensions of housing

55 x 55 x 250 mm   (2.16” x 2.16” x 9.84”)

Weight

570 g

Memory Data

No. of impacts in a series

Maximum 99

Memory capacity dependent on length of test series

Example

400 series with 10 values per series

Example

200 series with 20 values per series 

Electrical Data

Display

17 x 71 pixel, graphic

Accumulator duty

>5000 impacts between charges

Charger connection

USB type B (5V, 100 mA)

Environmental conditions

Operating temperature

0 to 50°C 

Storage temperature

-10 to 70°C

IP Classification

IP 54

8.  Maintenance and Support

Rebound hammers require periodic servicing and verification to provide reliable results.

Self test function

The RockSchmidt performs a number of self test checks and gives an alert when there is a problem.

These three alerts show the battery status:

 
Low battery

Battery full

Battery charging

Low impact energy: The impact spring is out of tolerance and should be replaced. It is 
recommended to return the RockSchmidt

 

to an authorized service centre.

Checking the calibration on the anvil

The impact spring is a consumable item that deteriorates over time. The RockSchmidt checks for 
correct impact energy with each and every impact, (see above). Nevertheless it is recommended to 
check the calibration on a regular basis. 
Set the hammer to single shot mode and perform at least three impacts before taking any readings. 

Perform a series of 10 impacts on the anvil (Part No. 341 10 400)
Check that the average R-value lies within the tolerance shown on the label. (e,g. 91+/-2 R for N-type).
If it is out of tolerance, check for smooth operation of the plunger. This problem can be solved by 
applying one or two drops, (no more), of light oil (e. g. sewing machine oil). If the plunger is dirty, 
wipe it with a soft cloth before applying the oil. After applying the oil, turn the plunger a few times 
and depress it a few times to spread the oil evenly. Wipe off any excess oil before use. If it is still out 
of tolerance it is likely that the impact spring needs replacing. 
Short term, use the correction factor method (see chapter 3.4) to continue testing.
Long term, return the hammer to a Proceq service centre for servicing.

Plunger replacement

Plungers may break when impacting on very hard materials or by impacting at an angle, i.e. not 
perpendicular to the surface. If you need to replace it, simply pull out the old plunger being careful 
not to lose the spring (see Note below) and replace it with a spare plunger, (Part No. 343 10 310).

NOTE! There is a small spring located inside the plunger. Take care not to lose this when 
removing the plunger for cleaning. Therefore hold the RockSchmidt with the plunger 
downward. Pull out the plunger and remove the spring. 

Support Concept

Proceq is committed to providing a complete support service for this instrument by means of our 
global service and support facilities. It is recommended that the user register the product on www.
proceq.com to obtain the latest on available updates and other valuable information.

Содержание 34310000

Страница 1: ...Made in Switzerland ROCK SCHMIDT OPERATING INSTRUCTIONS...

Страница 2: ...terface 8 2 5 Sample Preparation 9 2 6 Correct use of the Hammer 9 3 Optional Settings 10 3 1 Units 10 3 2 Statistics 10 3 3 Conversion Curve 11 3 4 Correction Factor 11 3 5 Checking the Current Setti...

Страница 3: ...ration and maintenance of the instrument and its components Unauthorized structural modifications to the instrument and its components Serious damage resulting from the effects of foreign bodies accid...

Страница 4: ...hut down automatically after 5 minutes of inactivity Standards and guidelines applied The following standards and guidelines have been applied in RockSchmidt for the procedure to determine the rebound...

Страница 5: ...en holder to prevent damage to the cores on impact Ideally it should have a semi circular groove with the same diameter as the core but it may also be a V shaped groove Please refer to the guidelines...

Страница 6: ...nd the range Once downloaded to a PC RockLink also calculates the median and the mode If a conversion curve has been selected tilting the instrument shows the UCS in the chosen unit Tilting again disp...

Страница 7: ...are not saved Select single shot mode Confirm or cancel the action 5 2 Measurement Series The number of impacts required for a valid rebound number depends on the statistics method selected Having sel...

Страница 8: ...lect review list Scroll in the list by ROLLING the hammer away from you or towards you The arrow indicates the current series Select summary for details of the series Select to delete the series Selec...

Страница 9: ...over time The RockSchmidt checks for correct impact energy with each and every impact see above Nevertheless it is recommended to check the calibration on a regular basis Set the hammer to single shot...

Страница 10: ...ton to establish the connection 9 1 Viewing the Data The data stored on your RockSchmidt will be displayed on the screen The test series is identified by the Impact counter value There is a Name colum...

Страница 11: ...exported in the format defined by the active Tab i e either in series or summary format To open the file in Excel locate the file right click on it and Open with Microsoft Excel The data will be opene...

Страница 12: ...eters and the hammer type RockSchmidt N for which it is applicable NOTE The curve may be exponential or a first second or third order polynomial ISRM recommend the use of exponential curves for UCS an...

Страница 13: ...07 info asia proceq com Proceq Rus LLC Ul Optikov 4 Korp 2 lit A Office 410 197374 St Petersburg Russia Phone Fax 7 812 448 3500 info russia proceq com Proceq Middle East P O Box 8365 SAIF Zone Sharja...

Отзывы: