PRO-790
7. Error Massage List
Follow the remedy precisely as shown in the following lists. If it is not possible to eliminate the problem, there is
the possibility of the splicer being faulty and the splicer may require service. Consult your nearest us with the
following information:
Model name of the splicer
Serial number of the splicer
Error message
Situation when the error occurs
Error Message
Reason
Remedy
Replace left fiber
Replace right fiber
Replace both fibers
The left fiber is set too far back.
The right fiber is set too far back.
The left or right fiber is set too back.
Reset, Moves left fiber forward
Reset, Moves right fiber forward
Reset, Moves left/right fiber all
forward
Reset, Moves left or right fiber
forward
Left cleave bad
Right cleave bad
Both cleave bad
Bad fiber end-face
Dust or dirt on the fiber surface.
“End-face angle” set up too strict
Dust or dirt on the objective lens or the
wind protector mirror.
Check the condition of fiber
cleaver. When the blade is worn,
rotate the blade.
Put“ End-face angle” loose to
suitable degree
Anew preparation fiber
Lean the lens or mirrors
Please close the wind
protector
Unable to start splicing when the wind
protector opens.
The splicer automatically starts
splicing after closing the wind
protector
The wind protector is opened during
splicing operation.
Press
reset after closing the
wind protector
Fusion
failure
The fiber stuff amount is insufficient.
Increase stuff amount in the
parameter setup menu
The pre-fuse power is too strong.
Minish pre-fuse power in the
parameter setup menu