13.
Modbus
13.1. Supported MODBUS features
106
The TPK series pressure dew-point monitor is equipped with a Modbus RTU interface. The
following points describe the MODBUS slave functionality.
Parameter
Value(s)
Device type
Slave
Baud rates
9600, 19200,
38400*
, 57600,
115200,230400, 460800, 921600 bps
MODBUS protocol
RTU over serial interface*
, ASCII over serial
interface
Data size
RTU-mode - 8*
ASCII mode - 7 bits
Stop bits
1
Parity
None*
, even, odd
Slave adress range
1⁄.247
* The colored items are factory settings
The MODBUS communication over serial line settings and slave address can be configured
over SENCON in our factory.
Physical Parameter
MODBUS-Address
Pressure Dewpoint / Frostpoint in
o
C
1600
System pressure in bar
576
ppm Volume concentration in ppm
1664
Absolute humidity in g/m
3
1920
Temperature in
o
C
1280 (Temperature is in Kelvin)
Relative humidity in %
1408