PRIZJet
Compressor nebuliser
27
8. Troubleshooting
Check the following if your device should fail during operation. You can
also refer to the pages of this Instruction manual for complete instructions.
Symptom
Cause
Remedy
Nothing
happens
when the
power switch
is pressed.
Is the power cord plugged into
the compressor and an electrical
outlet?
Check that plug is inserted into the
jack on rear side of compressor and
in an electrical outlet. Unplug, then
reinsert the plugs if necessary.
Is fuse on rear side of the
compressor ok?
Remove fuse cap on the rear side
of the compressor. Remove blown
fuse. Insert a new fuse in fuse holder.
Insert and lock fuse cap.
No
nebulisation
or low
nebulisation
rate, when
the power
is on.
Is there medication in the
medication tank?
Fill the medication tank.
Is air valve opened?
Press air valve against medication
tank.
Is there to much/little medication
in the medication tank?
Add the correct amount of
medication to the medication tank.
Is there vaporiser head missing
and not assembled correctly?
Attach the vaporiser head correctly.
Is the nebuliser kit assembled
correctly?
Attach the nebuliser kit correctly.
Is the nozzle blocked?
Make sure that nozzle is free of
blockages
Is the nebuliser kit tilted at a sharp
angle?
Make sure that the nebuliser kit is
not tilted at an angle more than 45
degrees.
Is the air tube connected
correctly?
Make sure that the air tube
is correctly connected to the
compressor and nebuliser.
Is the air tube folded or damaged?
Make sure that the air tube does not
contain kinks.
Is the air tube blocked?
Make sure that the air tube is free of
blockages?
Is the air filter dirty?
Replace the air filter with a new one.