10.0
Fault Finding
If a problem arises with the Hoist, the Table below will hopefully assist in determining the fault and what actions you can take. If
the fault cannot be found or the fault is found and the action guide does not provide a fix (e.g.
–
a damaged wire would need
replacement), contact your local Prism Medical UK authorized dealer immediately, an approved professional will be required to
repair the Hoist. Contact details can be found on the last page of this manual.
Fault
Action
The Handset has become disengaged
from the Hoist, or the Handset buttons
are not responding.
Refer to the section 4.3 ‘Connecting
the Handset to
The Hoist’.
If this does not correct the fault, then contact your local authorised dealer
immediately so the hoist can be checked to ensure proper continued operation.
The handset button command is
continuously activated
–
UP, DOWN, E-
LOWER.
Turn off the Hoist using the OFF cord by pulling it down. Contact your local
authorised dealer immediately so that the hoist can be checked to ensure proper
continued operation.
The carry bar of the Hoist does not
move UP or DOWN even when the
handset has been properly connected.
The indicator light on the control panel should be green and show that there is
power. If it is not, then pressing any coloured button on the handset to activate
the hoist and the indicator light should turn GREEN.
If the hoist still does not function, then the batteries may be low and require
charging. Refer to the section 5.5 'Charging the Hoist '. Charge the hoist for at
least one hour and then try to raise/lower the carry bar.
If none of these resolve the fault, DO NOT use the Hoist. Contact your local
authorised dealer immediately so that the hoist can be checked to ensure proper
continued operation.
The Hoist LED’s indicate there is power,
but the Hoist does not operate in the
DOWN direction.
A built-in detector checks the slackness of the lift tape. This may be sensitive.
Apply weight to the carry bar while pressing the DOWN button at the same time.
If this corrects the fault temporarily but not permanently then contact your local
authorised dealer so that the Hoist can be checked to ensure proper continued
operation
The red indicator light on the hoist turns
RED and/or a loud alarm sound is heard
when an individual is raised.
The batteries are low and require charging. Refer to section 5
.5 ‘Charging
the
Hoist’ and charge the hoist for at least one hour before trying to raise/lower the
carry bar.
If this does not correct the fault, then contact your local authorised dealer
immediately so that the hoist can be checked to ensure proper continued
operation.
One side of the lift tape is starting to
fray after continued use.
Contact your local authorised dealer immediately so the hoist can be checked to
ensure proper continued operation.
The hoist does not pass through a track
component such as a turntable or gate.
Refer to the User Manual of the specific piece of equipment in question.
If the recommended solution does not correct the fault, then contact your local
authorised dealer immediately so that the track component and hoist can be
checked to ensure proper continued operation.
Table 11
Содержание 108601
Страница 1: ...Doc No 999016 Revision of document G Revision Date 08 12 2021 CP Ceiling Track Hoist User Manual...
Страница 39: ...999016 Revision G P a g e 39 of 44...
Страница 40: ...999016 Revision G P a g e 40 of 44...
Страница 41: ...999016 Revision G P a g e 41 of 44...
Страница 42: ...999016 Revision G P a g e 42 of 44 User notes...
Страница 43: ...999016 Revision G P a g e 43 of 44 User notes...