Casters and Braking:
The FGSH-450 lift has two rear casters with brake. The rear casters can be braked for rotation, lateral move-
ment, and parking. To apply the brake, press the brake pedal down with your foot. To release the brake, press
the raised pedal towards the wheel.
During the lifting, the rear wheels should remain unlocked so that the lift will move to the patient’s centre of
gravity.
Application:
If the lift is used incorrectly any warranty or product liability might cease to be valid.
The FGSH-450 lift must only be used for person lifting and only for persons who, including the sling, do not
weigh more than the stated max. weight load. If the load exceeds the stated weight limit and if the lift is used
for lifting anything but persons, then any product liability that Prism Medical might have, in connection with
insurance / warranty / maintenance etc., will cease to be valid.
To avoid possible accidents and injury to persons being lifted, the floor lifts must only be operated as de-
scribed in the preceding pages.
Warning:
Locked wheels during lifting will increase the risk
of the lift tilting over.
Operation… continued