background image

 
 

Owner’s Manual 

 

Use and Care 
Trouble Shooting 
Warranty Information 

 
 

 
  FGSH-450  

by Prism Medical

 

Introduction ................................................................... 2 
 
Overview ....................................................................... 2 
 
Components of the Floor Lift  ........................................ 3 
 
Specifications ................................................................ 4 
 
Cautions ........................................................................ 5 
 
Shipping & Transportation Instruction  ........................... 6 
 
Assembly Instruction..…..…………………………………..7 
 
Operation 
      Leg Spreading.....……………………………………...12 
      Raising and Lowering Boom………………………….14 
      Forward movement….………………………………...15 
      Turning..………………………………………………...15 
      Casters and Braking……….……………………….....17 
           
Basics in transferring an individual…………………...…..17 
 
General care, inspection and maintenance....…………..20 
 
Points of Attachment - Checklist …………….…………...24 
 
Service record history……………………………..…….....25 
 
Warranty……………………………………………..………29 

Содержание FGSH-450

Страница 1: ...ifications 4 Cautions 5 Shipping Transportation Instruction 6 Assembly Instruction 7 Operation Leg Spreading 12 Raising and Lowering Boom 14 Forward movement 15 Turning 15 Casters and Braking 17 Basic...

Страница 2: ...rofessionals to transfer cli ents The floor lift makes it possible to transfer mobility impaired individuals with minimal strain or risk to the caregiver while providing complete safety dignity and co...

Страница 3: ...uator Carry Bar Boom Rear Caster w Brake Leg Front Caster Push Handle Pump Handle The following components are included with your new FGSH 450 Hydraulic Steel Floor Lift system FGSH 450 Hydraulic Stee...

Страница 4: ...17 inches 43 2 cm Height to top of legs 6 5 inches 16 5 cm Clearance from bottom of legs to floor 5 25 inches 13 3 cm Overall length 46 75 inches 118 75 cm Distance inside the legs min 16 75 inches 4...

Страница 5: ...he Hydraulic Pump as this will VOID THE WARRANTY In facilities where more than one operator will be responsible for using the FGSH 450 Hydraulic Steel Floor Lift and sling s it is imperative that all...

Страница 6: ...and remove loose packing materi al and following items and place on the floor taking care to protect the finish from damage FGSH 450 Base Assembly FGSH 450 Mast and Boom Assembly with carry bar User M...

Страница 7: ...these notch at the bottom of mast assem bly with the tabs inside the tube of base assembly Figure 4 as shown in Figure 5 3 Insert the mast assembly into the tube of base assembly and onto the TABS Ca...

Страница 8: ...he base assembly and securely tighten as shown in Figure 7 and 8 NOTE If the mast does not turn then the mast is centered and locked in place Figure 6 Figure 8 Mast Locking Screw Figure 7 Warning The...

Страница 9: ...ll pivot easily if the mounting hardware is aligned properly To check for correct alignment of the hardware use the Allen key and turn the shoulder bolt by hand The shoulder bolt should rotate without...

Страница 10: ...14 Insert the Base Shifter Handle into the Cam lock Assembly at the rear of the base as shown in Figure 15 2 Tighten the thumb screw to secure the shifter handle in place as shown in Figure 16 Figure...

Страница 11: ...ry bar pin and washers are completely pushed up through the Carry Bar Pad The carry bar has hooks at both ends to attach the sling straps that supports the patient during the lift The carry bar pad sh...

Страница 12: ...hole as shown in Figure 18 3 Turn the handle CLOCKWISE until you are able to secure the lock pin into the opposite mounting hole to fully open the legs of the base as shown in Figure 19 NOTE The lock...

Страница 13: ...DO NOT use the patient lift until the lock pin is properly seated and LOCKED in place Failure to comply with this caution could result in serious injury to the indi vidual being lifted and or damage...

Страница 14: ...the Boom The control valve must be in the OPEN position control valve positioned away from pump handle to lower the boom and the patient as shown in Figure 24 The rate of descent can be controlled by...

Страница 15: ...foot Turning When turning the lift stand on along one side With one hand pull gently on the push handle and with the other hand push gently away on the lifting arm In this way the lift will rotate aro...

Страница 16: ...ny warranty or product liability might cease to be valid The FGSH 450 lift must only be used for person lifting and only for persons who including the sling do not weigh more than the stated max weigh...

Страница 17: ...with the sling that was purchased on how to properly outfit an individual with a sling Step 1 Using the Push Handle push the patient lift underneath of the bed Caution Always use extreme care when mo...

Страница 18: ...ps to chain hooks Failure to comply with this caution could result in serious injury to the individual being lifted and or damage to the lift Step 5 The individual may now be raised by pumping the lif...

Страница 19: ...desired position DO NOT lock the rear casters when lifting the individual LOCKED Rear casters during the lifting will increases the risk of the lift tilt over Step 7 Slowly open the control valve to l...

Страница 20: ...ely HYDRAULIC PUMP All parts of the Hydraulic Pump are precision machined then carefully assembled and tested to ensure relia ble service The pump assembly is completely enclosed and sealed with Neopr...

Страница 21: ...ivot for slings for signs of crack ing fraying deformation or deterioration Replace any defective parts immediately and ensure that the lift is not used until repairs are made Check that all labels ar...

Страница 22: ...page fail the inspection do not use the lift Contact Prism Medical Systems or your local qualified service technician for service Complete the inspection as per check list on next page Check and lubri...

Страница 23: ...r engaged X X X Check roll pin for wear X X X MAST Mast should lock securely when assembled X X X Inspect for bends or deflections X X X BOOM Check all hardware and carry bar supports X X X Inspect fo...

Страница 24: ...chment Use the described tools to tighten the bolts on a monthly basis in the areas shown 1 2 Wrench 3 16 Allen Key Hydraulic Pump mounting Mast locking screw 1 2 Wrench 3 16 Allen Key Carry bar mount...

Страница 25: ...plete the following section on Purchase and Service Information as soon as this equipment is purchased Use the service record history to record to any completed service and repairs Ensure that the ser...

Страница 26: ...______________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Serv...

Страница 27: ...______________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Serv...

Страница 28: ...______________________ Printed Name Signature Company _____________________________________________________________ Remarks Action Taken Date _______________________ Time ________________________ Serv...

Страница 29: ...ty shall be void if the equipment is not serviced by Prism Medical or its authorized service agents in accordance with the manufacturer s recommendations or if any unauthorized person carries out work...

Отзывы: