1. Introduction 

The PriorLux POL upright compound polarised light microscope is a high quality instrument 

equipped with high resolution, chromatically corrected strain-free optics for excellent image 

quality in polarised light applications. The robust construction and hard wearing materials 

ensure long lasting and trouble free operation. The instrument can be used with a number of 

supplied accessories including ¼  and   compensation plates, quartz wedge, attachable 

mechanical XY stage and eyepiece graticule with stage micrometer. 

 

For documentation the instrument is available with a trinocular head which permits mounting 

of a video or digital camera system. 

 

2. Unpacking 

The PriorLux POL is shipped in protective bags within a pre-formed container. Each 

component should be carefully unpacked and checked, cutting rather than tearing the plastic 

bags. The head (binocular or trinocular) should be fitted to the dovetail on top of the stand 

and locked in place with the head locking screw. The eyepieces then just drop into the 

eyepiece tubes at the front of the head, these should be pushed in as far as they will go. 

 

Remove each objective lens from its protective “pot” and screw into the free positions on the 

nosepiece. Move the stage and the condenser carrier to the highest position and insert the 

condenser from underneath with the diaphragm control facing to the front. This is locked in 

position with the clamping screw, which is fitted from the side.

Содержание Lux POL

Страница 1: ...Operating Instructions PriorLux POL Advanced Polarising Microscope...

Страница 2: ......

Страница 3: ...tion 6 8 Cleaning Objectives 7 9 Centring the Objectives 7 10 Attaching the Mechanical XY Stage 8 11 Observations Between Crossed Polars 8 12 Conoscopic Observations 8 13 Bertrand Lens 9 14 Use of Com...

Страница 4: ...nting of a video or digital camera system 2 Unpacking The PriorLux POL is shipped in protective bags within a pre formed container Each component should be carefully unpacked and checked cutting rathe...

Страница 5: ...1 degree increments Focusing Mechanism Co axial fine and coarse adjustment with tension control and focus stop Viewing Head Interpupillary distance 55 75mm Kohler Illumination 12V 30W halogen lamp wit...

Страница 6: ...4 Prior Scientific 4 Component Parts...

Страница 7: ...hing on the power reduce the lamp intensity control to its minimum setting After switching on the intensity should be increased to a comfortable level This procedure should be reversed before switchin...

Страница 8: ...ld to attain perfect centration Figure 1 Pictures C D E Fully open the field diaphragm F Fully open the condenser diaphragm and then slowly close it until you see the contrast within the image increas...

Страница 9: ...erefore centration of this particular lens is not user adjustable All other powers are mounted in positions with adjustment screws fitted into the nosepiece above to enable the end user to make adjust...

Страница 10: ...cally when the back focal plane of the objective is observed using the Bertrand lens In petrology birefringent specimens can produce interference patterns in this plane that are of considerable import...

Страница 11: ...rs will be observed The retardation caused by the specimen can be assessed by its compensation of an order of colour of the wedge The Prior quartz wedges are mounted length fast and when the interfere...

Страница 12: ...on of the selected video or digital camera are supplied with the camera Assembly Video Cameras A Screw the c mount adapter part no WXCM1 1 0x or WXCM050 0 5x to the video camera B Loosen the knurled s...

Страница 13: ...PriorLux POL 11...

Страница 14: ...its back to gain access to the base plate D Loosen the screw which holds the lamp cover to the base plate E Open the lamp cover to expose the bulb holder F Remove the old bulb and replace it with the...

Страница 15: ...PriorLux POL 13 17 Fuse Location The fuse is located on the base towards the front right corner of the instrument 18 Spare Parts W3257 Spare bulb 12V 30W Halogen W335 Fuse T1 25A...

Страница 16: ...bu lba cces sdoor onthe under s i de o ft he i n s t rument Warn ing H i ghvo l t a ge d i s connec tpowersupp l ybef o rechang ingthe bu lb This symbol is located next to the bulb access door on the...

Страница 17: ...se part 1 General requirements EN61326 1997 A1 A2 A3 Electrical equipment for measurement control and laboratory use EMC requirements Class B emmisions EN61326 1997 A1 A2 A3 Electrical equipment for m...