16
PI-MAX
®
3 System Manual
Version 1.A
Cleaning and Maintenance
Cleaning the Camera
Although there is no periodic maintenance that
must
be performed on the PI-MAX3
Camera, users are advised to wipe it down with a clean damp cloth from time to time.
Note:
The cloth should be just damp enough to pick up dust – not wet.
This operation should only be done on the external surfaces and with all covers secured.
In dampening the cloth, use clean water only. No soap, solvents or abrasives should be
used. Not only are they not required, but they could damage the finish of the surfaces on
which they are used.
Cleaning Optical Surfaces
– Unigen
TM
II Coating
For greater sensitivity, the PI-MAX3 camera with Unigen II coating comes without a
protective window over the intensifier. Do not scratch the image intensifier’s coated
surface. To remove any lint or dust use very clean dry nitrogen with < 5 PSI pressure. If
you have any questions please contact your sales contact or the factory.
Cleaning Optical Surfaces
– Protective Window
Optical surfaces may need to be cleaned due to the accumulation of atmospheric dust. We
advise that the
drag-wipe
technique be used. This involves dipping a clean cellulose lens
tissue into clean anhydrous methanol, and then dragging the dampened tissue over the optical
surface to be cleaned. Do not allow any other material to touch the optical surfaces.
Flushing and Refilling the CCD Chamber
Under normal conditions the CCD chamber is sealed and backfilled so there is no danger
of damage due to condensation.
Operating an Intensified PI-MAX3
that is no longer backfilled with dry air or dry
nitrogen may result in condensation on the array that could cause irreversible damage.
Such damage would not be covered by the Warranty.
Before a PI-MAX3 camera leaves the factory, its CCD chamber is backfilled with clean
dry air or dry nitrogen. For proper operation it is essential that the integrity of the front
enclosure be maintained
.
In normal operation, the CCD chamber should remain sealed for the life of the detector
and should require no maintenance to assure integrity. If it should ever happen that the
CCD chamber becomes unsealed, contact the factory and arrange to return the detector to
the factory where it can be properly flushed, backfilled and resealed again.
See page 146
for contact information.
Repairs
Save the original packing materials. Because the PI-MAX3 system contains no user-
serviceable parts, repairs must be done by Princeton Instruments. Should your system
need repair, contact Princeton Instruments customer support for instructions (telephone,
e-mail, and address information are provided on page 146 of this manual).
Use the original packing materials whenever shipping the system or system components.
WARNING
!
WARNING
!
Содержание PI-MAX 3 System
Страница 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Страница 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Страница 60: ...60 PI MAX 3 System Manual Version 1 A Figure 24 Experiment is Master Clock Hardware Setup and Timing Diagram ...
Страница 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Страница 111: ...Appendix B Outline Drawings 111 Figure 69 Outline Drawing PI MAX3 with Spectroscopy mount Adapter ...
Страница 112: ...112 PI MAX 3 System Manual Version 1 A PI MAX3 Power Supply Figure 70 Outline Drawing PI MAX3 Power Supply ...
Страница 113: ...Appendix B Outline Drawings 113 CoolCUBEII Circulator Figure 71 Outline Drawing CoolCUBEII Circulator ...
Страница 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 142: ...142 PI MAX 3 System Manual Version 1 A ...