116
PI-MAX
®
3 System Manual
Version 1.A
Main tab page
(Experiment Setup on Acquisition menu)
Figure 73. Main tab page with PI-MAX3
camera
The Experiment Setup
Main
tab page is used to specify the most essential parameters.
These include the Exposure Time, Number of Images/Spectra, Number of Accumulations
and choice of Full Chip or Region of Interest readout. Some of the selections on this page
only apply for a PI-MAX3
camera. With other cameras these selections are not displayed.
Selections unique to the PI-MAX3
include.
Intensifier:
Gain:
Sets the intensifier gain when operating with a PI-MAX3
intensified
camera. The range is in arbitrary units from 0 to 100. The default setting will
give good results in many applications.
Gate Mode:
Selecting this radio button puts PI-MAX3
in Gate mode, in which
the photo-cathode is biased on by pulses generated by the Timing Generator.
Safe Mode:
Selecting this radio button puts PI-MAX3
in the Safe mode, in
which the photocathode is continuously biased off until
Gate Mode
is selected.
Note that Safe mode is the turn-on default.
In a system having the PI-MAX3
camera, when the software is booted, a dialog box appears that allows you to
change the PI-MAX3
start-up gate mode.
Содержание PI-MAX 3 System
Страница 1: ...4411 0129 Version 1 A September 8 2010 4411 0069 ...
Страница 32: ...32 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 52: ...52 PI MAX 3 System Manual Version 1 A Figure 20 Safe Mode and Fast Mode Operation ...
Страница 60: ...60 PI MAX 3 System Manual Version 1 A Figure 24 Experiment is Master Clock Hardware Setup and Timing Diagram ...
Страница 72: ...72 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 90: ...90 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 100: ...100 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 110: ...110 PI MAX 3 System Manual Version 1 A Figure 68 Outline Drawing PI MAX3 with F mount Adapter ...
Страница 111: ...Appendix B Outline Drawings 111 Figure 69 Outline Drawing PI MAX3 with Spectroscopy mount Adapter ...
Страница 112: ...112 PI MAX 3 System Manual Version 1 A PI MAX3 Power Supply Figure 70 Outline Drawing PI MAX3 Power Supply ...
Страница 113: ...Appendix B Outline Drawings 113 CoolCUBEII Circulator Figure 71 Outline Drawing CoolCUBEII Circulator ...
Страница 114: ...114 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 120: ...120 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 140: ...140 PI MAX 3 System Manual Version 1 A This page intentionally left blank ...
Страница 142: ...142 PI MAX 3 System Manual Version 1 A ...