background image

     

DOC-040-2350 PCB901 Rotating Half Rail Instructions| REV 001 

 

The information in this document is subject to change without notice. 

Page5 

 

 WARNING 

Bed should be at lowest suitable height for entry and exit. 

 

 ENTRAPMENT WARNING 

Accurate assessment of the patient and monitoring of correct maintenance and use of 

equipment are required to prevent entrapment. For additional information on product 

and safety issues for bed frames and rails review the manual specific to the products 

you are using.  If bed frames have been serviced or any other adjustments have been 

made, make sure all parts are securely back in place before operating the bed frame.  

 

 ENTRAPMENT WARNING 

The risks associated with these devices used in conjunction with specialty mattresses and 

mattress overlays have not been evaluated.  When these specialty mattresses or mattress 

overlays are used the end use facility must take steps to ensure that the therapeutic benefit 

outweighs the risk of entrapment. 

 ASSIST RAIL WARNING 

Other manufacturers assist rails/bars may not be compatible and can lead to 

entrapment issues or harm to patients.  Only compatible Primus Medical assist bars and 

rails may be used on this bed.  Make sure mattress is the correct size for bed frame and 

the assist bars are secured to frame to decrease the risk of entrapment. 

 

 

Installation: 

1.

 

Remove the half rails from the packaging. 

2.

 

Remove the 2 knobs from the mounting bolts for one half rail. 

 

3.

 

Install the left or right side half rail by lowering the half rail over the top of the bed deck 

lining up the holes in the half rail mounting bracket with the holes in the square tube on 

the deck. 

NOTE:  There are alignment pins that must be paid attention to.  These pins help with the 

identification of the appropriate mounting positions. 

Содержание PRIMECARE PCB901

Страница 1: ...DOC 040 2350 PCB901 Rotating Half Rail Instructions REV 001 The information in this document is subject to change without notice Page1 ...

Страница 2: ...ure 10 C 50 C 14 F 122 F environment Relative humidity range 10 93 Maximum patient weight safe working load This device is not intended to support the entire weight of a patient This device has been tested to allow a downward load of 750 N 168 lbf and a side load of 500 N 112 lbf Description of device function This device has a single upright locked position When the upright locked position is not...

Страница 3: ... pending hazards that if not avoided will result in immediate serious personal injury or death in addition to damage of the equipment WARNING Information that appears under the WARNING description concerns the protection of personnel from possible hazards that can result in injury or death in addition to damage of the equipment CAUTION Information that appears under the CAUTION description concern...

Страница 4: ... bed frame WARNING This device rotates When routing cables from other equipment in the medical bed precautions shall be taken to avoid squeezing those between parts of the medical bed WARNING No modification of this equipment is allowed WARNING Do NOT use assist bars as handles for moving the bed WARNING Keep all moving parts free of obstructions WARNING NEVER allow anyone under the bed at any tim...

Страница 5: ...en evaluated When these specialty mattresses or mattress overlays are used the end use facility must take steps to ensure that the therapeutic benefit outweighs the risk of entrapment ASSIST RAIL WARNING Other manufacturers assist rails bars may not be compatible and can lead to entrapment issues or harm to patients Only compatible Primus Medical assist bars and rails may be used on this bed Make ...

Страница 6: ...notice Page6 4 Insert the 3 8 x2 5 square neck bolts from the outside of the bed frame and install the knobs on the ends of the bolts WARNING The knobs must be installed tightly to ensure proper mounting of the device 5 Repeat the procedure for the opposite side Assist bar mounting positions and orientations Head deck only ...

Страница 7: ...HIS DEVICE IS NOT COMPATIBLE WITH WASHDOWN TUNNELS Inspections WARNING Failure to take care of your device may decrease the life of your product and increase maintenance Always inspect and service the device at the required intervals INITIAL INSPECTION Inspect All Components Check components for obvious damage Verify proper function Quarterly Maintenance Check Inspect all welds for cracks Ensure p...

Страница 8: ...DOC 040 2350 PCB901 Rotating Half Rail Instructions REV 001 The information in this document is subject to change without notice Page8 ...

Отзывы: