With the button
Copy
the measurement settings (window size and position, power and amplification) can
be copied from the previous measuring plane.
By means of the option
Averaging
the average of the results of up to 50 single measurements per each
plane is determined. There are different analysis algorithms available:
Selection
Function
Average
determines the average value of the distributions measured
Max. pixel
determines the pointwise maxima of the distributions measured
Max. trace
determines the maximum traces of the distributions measured
The selection
Max. pixel
,
Max. trace
are especially helpful when it comes to pulsed laser radiation. The radii
determined in case of
Max. pixel
are not always reliable due to zero point uncertainties.
During a measurement, the status of the measurement system is constantly displayed. These are:
•
the current measuring plane
•
the run of the reference cycle
•
positioning the measuring head
•
the measurement
•
the data transmission – the progress is shown by means of the bar display
By means of the button
Stop
you can cancel a running measurement which also ends the monitor operation.
Please click on the
Reset
button afterwards.
With the button
Stop Motor
the rotation of the measuring tip is stopped after the completion of the current
measurement. Please click the Reset button afterwards.
CAUTION
Danger of injury due to rotating parts
The measuring tips of the BeamMonitor BM+ keep rotating for a short period of time after
the supply voltage was turned off.
Do not reach into the entrance of the measuring device and do not hold any items into it, as
long as the measuring tip is rotating.