![Primes BM+ 100 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/primes/bm-100/bm-100_operating-manual_1601731014.webp)
PRIMES
BeamMonitor BM+ with LDS 2.98
14
Revision 01/2018 EN
7
Installation
7.1
Preparation
The measuring device must be assembled stably and must be affixed by means of screws (see chapterl
7.4 on page 15).
DANGER
Fire hazard; Damaging/Destruction of the device
After passing the device, the laser beam has to be absorbed completely. Fire bricks or other
partly-absorbing surfaces are not suitable!
Please use an adequate absorber, e.g. the PRIMES PowerMonitor.
7.2
Installation position
The BeamMonitor BM+ can also be mounted horizontally or vertically.
Mounting surface
BeamMonitor BM+
Mounting surf
ace
BeamMonitor BM+
Mounting surface
BeamMonitor BM+
Mounting surf
ace
BeamMonitor BM+
Horizontal mounting
Vertikal mounting
Beam incidence
Fig. 7.1:
Mounting options for the BeamMonitor BM+
7.3
Alignment
The BeamMonitor BM+ has to be positioned correctly and a solid assembly needs to be ensured. The beam
has to hit the measuring aperture perpendicular and centrally.
DANGER
Serious eye or skin injury due to laser radiation
If the proportion of the laser beam diameter with regard to the diameter of the aperture is
too large, scattered radiation could occur during the measurement operation.
The laser beam diameter must not exceed 0.7 times the aperture diameter.
Especially in case of a high beam quality, we would recommend to remain below 0.6. Otherwise, a falsification
of the measuring results due to the cutting off of border fields is to be expected. Especially when it comes
to the determination of radii according to the 2nd-moments-method, problems are possible. Moreover, there
is a danger of a heating up of the housing.