Description of device
METRAX GmbH
DefiMonitor XD 21324 / GB / K05
29 / 116
Display.
Meaning
Action to be taken
PRIMEDIC™ AkuPak
sufficient capacity
Device ready to use.
Battery low
You can use the device, but
the PRIMEDIC™ AkuPak may
need to be charged or
replaced.
Symbol also appears if no
power module is inserted!
Insert power module.
Symbol also appears event
hough use-by-date of power
module has passed.
Check use-by-date,
if necessary replace power
pack.
Battery symbol flashes during operation
Internal buffer battery empty
(The device is still
operational!)
Send the device to the dealer
for the replacement of the
internal buffer battery
PRIMEDIC™ AkuPak
sufficient capacity
Carry out extensive self-test
by reinserting PRIMEDIC™
AkuPak or switching the
device on again
Device defective
Have the device repaired by a
dealer.
Battery fully discharged
Charge battery
Device defective
Carry out extensive self-test
by reinserting PRIMEDIC™
AkuPak or switching the
device on again
Have device repaired by
authorised dealer