Jet1 HD/Rev H/Feb03
www.pridemobility.com
15
I I I . E M I / R F I
Are all power chairs susceptible to EMI/RFI?
Each make and model of chair differs in its ability to resist electromagnetic interference. That is, each has a
particular level of immunity to interference, measured in volts per meter (Vm). A higher immunity level offers
greater protection. In other words, a power chair with a high immunity level is less likely to be affected by a strong
radio source than one with a low immunity level.
What is the FDA doing about the problem?
The FDA has written to the manufacturers of power chairs and requested that they test their new products to be
sure that they provide a reasonable degree of immunity against EMI/RFI. The letter states that power chairs should
have an immunity of at least 20 V/m. This provides a reasonable degree of protection against the common sources
of EMI/RFI.
The FDA has also requested that these manufacturers clearly label their products with the immunity level, or state
that the immunity level is not known. The labeling and informational material supplied with the power chair must
explain what the immunity level means, and the labeling or informational material must warn users about the possi-
bility of EMI/RFI and how to avoid it. In addition, the FDA has recommended that manufacturers establish an
educational program to inform users of power chairs, and their caregivers, about the problems associated with
EMI/RFI, and the actions they can take to minimize the risk.
What can I do to find out if my power chair is likely to be affected by EMI/RFI?
If youve had your power chair for a long time and havent experienced any unintended motion, it is not likely that
you will have problems in the future. But it is always possible that problems could arise if you are close to a source
of radio waves. Therefore, it is very important to be alert to this possibility. The Jet meets or exceeds an immunity
level of at least 20 V/m.
What can I do to reduce the risk that my power chair could be affected by EMI/RFI?
Here are some precautions you can take:
n
Do not turn on or use hand-held personal communications devices, such as citizens band (CB) radios and
cellular phones, while the power chair is on.
n
Be aware of nearby transmitters, such as radio or TV stations and hand-held or mobile two-way radios, and
try not to come close to them. For example, if you are in a power chair with an immunity level of 20 V/m, you
should remain at least three feet from a hand-held two-way radio and ten feet from a mobile two-way radio.
n
Be aware that adding accessories and/or components, or modifying the power chair may make it more sus-
ceptible to interference from radio wave sources. There is no easy way to evaluate their effect upon the overall
immunity of the power chair.
What should I do if my power chair moves unexpectedly?
If unintended motion or brake release occurs, turn the power chair off as soon as it is safe to do so. Call Pride at
800-424-8205 to report the incident.