If you are using a numeric keyboard connected to the PCON-Y18-LP Type A USB
(see section
1.5
– Parts Identification), you can increment and decrement the set point
by pressing the “+” and “-” buttons in the keyboard.
Fig. 25
- NUDGE Function
2.1.4. Vent Mode
WARNING!
The calibrator discharges gas to the atmosphere through the vent port.
Leave this port free.
WARNING!
Before connecting the instrument under test to the output port, make a
VENT operation to discharge any gas inside the controller and go to
MEASURE mode.
In this mode, the calibrator turns the pressure control off and vents the pressure
inside the control module through the
VENT
valve.