PRESYS
Instruments
DMY-2030-TOT
Light
Maintenance
Page 37
Calibration of voltage input (0 to 5V)
In a 0 to 5V voltage input calibration connect a voltage voltage source to
terminals 2(+) and 3(-). It is required 6 voltage references listed in table 7.
Reference
Mnemonic
0.0000V
C. 0V
1.0000V
C. 1V
2.0000V
C. 2V
3.0000V
C. 3V
4.0000V
C. 4V
5.0000V
C. 5V
Table 7 - References for 0 to 5V input calibration
Calibration of voltage input (0 to 10V)
In a 0 to 10V voltage input calibration connect a voltage source to terminals
1(+) and 3(-). It is required 6 voltage references listed in table 8.
Reference
Mnemonic
0.0000V
C. 0V
2.0000V
C. 2V
4.0000V
C. 4V
6.0000V
C. 6V
8.0000V
C. 8V
10.0000V
C.10V
Table 8 - References for 0 to 10V input calibration
Calibration of current input (0 to 20mA)
In a 0 to 20mA current input calibration connect a current source to terminals
1(+) and 3(-). It is required 6 current references listed in table 9.
Reference
Mnemonic
0.000 mA
C. 0nA
4.000 mA
C. 4nA
8.000 mA
C. 8nA
12.000 mA
C.12nA
16.000 mA
C.16nA
20.000 mA
C.20nA
Table 9 - References for 0 to 20mA input calibration
presys