6. Perform Standard Strip test to assure that Meter
is working correctly
(see Sec. 3)
. If result is
within range of numbers printed on Standard
Strip label, go to Step 7. If result is NOT within
range printed on label, clean Test Window
(see
Sec. 7)
. After replacing Test Strip Platform,
perform another Standard Strip test. If second
test result is within range printed on label, go to
Step 7. If second test result is not within range,
call for assistance.
7. Perform Control test to assure that System
(Meter, Test Strips, and testing technique) is
working properly
(see Sec. 3)
.
8. If both Steps 6 and 7 give acceptable results, then
blood glucose results are correct.
9. If you still feel that your results are not correct,
contact your Diabetes Healthcare Professional,
or call Technical Service for assistance.
10.If Meter displays error message, please refer to
Section 9 DISPLAY MESSAGES
, for further
instructions.
11.If Meter beeps and flashes “Low Battery”
symbol, and then shuts off, battery is too low to
test. Replace battery
(see Sec. 8)
.
36