Once the sealing bag has been sealed it can be cut and removed from the front 
tray ready for the instruments to be placed inside. 

 

Once the sealing phase ends the LED will turn red and the buzzer will sound to 
signal that the sealing bar should be re-opened. If the lever is left in the down 
position for more than 10 seconds, the heating element will turn itself off 
automatically to protect the sealing bag and the components of the pouch sealer.  

After each sealing cycle the pouch sealer will check its temperature and if 
necessary it will activate the heater to keep the heating element at the required 
sealing temperature.  

Remove the packet created and put in the object to be sealed. Insert the open 
end of the packet into the front slot of the packet tray with the clear plastic film 
facing up. 

Warning: No object or instrument should be between the sealing bar and 

cutting tool as this could damage the heating element. Do not operate 

the lever without first inserting sealing bags. 

Содержание Pouch Sealer

Страница 1: ...Instructions for use Pouch Sealer...

Страница 2: ...cal co uk The Prestige Medical Customer Service Team is available to provide advice and assistance during normal office hours To avoid delays when making contact please have the unit s Model and Seria...

Страница 3: ...1 INTRODUCTION 4 FORWARD 5 PRESTIGE MEDICAL POUCH SEALER 5 SPECIFICATION 6 INSTRUMENTS 7 2 INSTALLATION 8 GENERAL CONDITIONS 8 WALL MOUNTING 9 3 OPERATION 10 SETUP 10 SEALING 10 4 USER MAINTENANCE 13...

Страница 4: ...oads are kept within a water tight and air tight seal for effective autoclaving and sterilization procedure They are suitable for mainly dentists beauticians podiatrists and in certain medical or labo...

Страница 5: ...with the operating conditions Such instructions are confidential and the customer is liable not to disclose any information to third parties Such operating instructions and its annexes shall not be c...

Страница 6: ...ions Width 470 mm Height 300 mm Depth 270 mm Packaging dimensions Width 480 mm Height 310 mm Depth 280 mm Reference standards The unit was designed in compliance with the following standards 73 23 CE...

Страница 7: ...otherwise switch the unit off Steady red light Sealing has finished where bar has been closed for 3 seconds Re open sealing bar by raising lever Red flashing light Sealing bar broken Replace the seali...

Страница 8: ...laboratory dentist or similar surgery where access is only allowed by authorised personnel It must not be installed Near a heat source In an area containing volatile or explosive substances Below or...

Страница 9: ...quired Repeat this process for the remaining hole to support the bracket Check the fixings are secure before mounting the unit It is advisable to use open hook expansion bolts to allow for easy instal...

Страница 10: ...and the heating element will be at its required temperature the pouch sealer is now ready for use Please Note If the lever is activated before the pouch sealer has reached the required temperature the...

Страница 11: ...off automatically to protect the sealing bag and the components of the pouch sealer After each sealing cycle the pouch sealer will check its temperature and if necessary it will activate the heater t...

Страница 12: ...uch sealer on and does not use if for at least 30 minutes it will go into ECO mode the sealing bar switches off to prevent damage and energy waste When it is in this mode the LED will flash green at a...

Страница 13: ...he maintenance must immediately report any anomalies scratches fraying etc Prestige Medical will accept no responsibility when procedures recommended by our service engineers are not heeded All mainte...

Страница 14: ...cted Replacing the blade Check and or replace the blade annually or if the blade has become blunt and cannot perform an easy cut DISPOSAL AND SCRAPPING Please dispose of its different parts in accorda...

Страница 15: ...to persons Do Not Disassemble the pouch sealer This should only be undertaken by an authorised service engineer Clean with abrasive or corrosive chemicals Immerse the pouch sealer in water Use the po...

Страница 16: ...s without notice Paper Please recycle Prestige Medical Limited East House Duttons Way Shadsworth Business Park Blackburn BB1 2QR England Registered in England Reg No 2826793 Vat No 633707835 Registere...

Отзывы: