211846B
Please read these instructions before using the
autoclave.
Keep these “Instructions for use” in a safe place
close by the unit for future reference.
Service Helpline: Telephone: 01254 844 116
e.mail: [email protected]
The Prestige Medical Customer Service Team is available
to provide advice and assistance during normal office
hours. To avoid delays when making contact, please have
the unit’s Model and Serial Numbers at hand.
For additional information visit
www.prestigemedical.co.uk
UK Customers
Prestige Medical Limited
East House, Duttons Way,
Shadsworth Business Park,
Blackburn BB1 2QR
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606
www.prestigemedical.co.uk
[email protected]
Contents.
Section
Page
1 Introduction
3
2
Operating symbols, displays
and controls
3
3 Operation
4
4
Do’s and Dont’s
5
5
Trouble shooting
6
6
Technical specification
7
7
Additional information
8
Instructions for use
Model no.
Date of purchase
Serial no.
Overseas
Customers
Contact your local
distributor. In case of
doubt contact Prestige
Medical.
0086
Prestige Medical Limited
East House, Duttons Way,
Shadsworth Business Park,
Blackburn BB1 2QR
Tel : +44 (0) 1254 682 622
Fax : +44 (0) 1254 682 606
www.prestigemedical.co.uk
[email protected]
Registered in England
Reg No. 2826793
IMPORTANT OPERATING INFORMATION
CAUTION!
Before using or maintaining the Prestige Medical Classic
Media autoclave ensure that you read and fully understand
the operating instructions paying particular attention to
information held within this page.
This model is designed to operate at 121 +2ºC
for 22 minutes, for the sterilization of culture
media. MAXIMUM LOAD 500ml. Please note
that the Classic Media should not be used to
sterilize wrapped instruments or porous loads. This cycle is
not recommended for the sterilization of medical or dental
instruments. TST Strips are not supplied with this model.
When sterilizing media, to ensure the safe operation of the
autoclave and to avoid damage to the media, DO NOT
manually vent the pressure by operating the pressure
release valve at the end of the cycle. As an added safety
feature , the lid interlock has been set to release only when
the temperature of the chamber contents has dropped to
80ºC or below.
2