8. Trouble shooting   

                                                         

MRJ

 

 

 

Symptom Cause 

Remedy 

Mechanics

 

The cover glass faces down. 

Put the cover glass to face up. 

(1)Can not get the 
objective focused. 

The cover glass is not standard.   

Use a standard cover glass with 
thickness 0.17mm. 

The cover glass faces down. 

Put the cover glass to face up. 

(2) The objective 
touches the cover 
glass while turning 
the nosepiece.   

The cover glass is not standard. 

Use a standard cover glass with 
thickness 0.17mm. 

(3)Coarse focusing 
knob is too tight. 

Tension knob is too tight. 

Loosen a little. 

(4)Stage declines 
itself and can’t stay in 
the focus plane. 

Tension knob is too loose. Tighten 

little. 

(5)Coarse focusing 
knob can’t rise. 

The limit stop knob is locked. 

Loosen the knob. 

(6)Coarse focusing 
knob can’t decline. 

The base of the condenser is too low. 

Raise the base. 

The slide is not fixed correctly. 

Adjust it correctly. 

(7)Can not move the 
slide smoothly. 

The movable specimen holder is not fixed 
properly. 

Adjust it correctly. 

(8)The image moves 
obviously when 
touching the stage. 

The stage is fastened incorrectly. 

Fasten the stage correctly. 

Electrical 

No power supply. 

Check the connection of the 
power cable. 

The bulb is not inserted correctly. 

Insert it correctly. 

(1)The bulb does not 
work. 

The bulb burnt out. 

Replace it. 

(2)The bulb burnt out 
usually. 

Use a wrong bulb. 

Replace with a correct one. 

Use a wrong bulb. 

Replace with a correct one. 

(3)The field of view is 
not bright enough. 

The use of light adjustment knob is wrong.   

Adjust correctly. 

The bulb will burn out soon. 

Replace with a new one. 

(4)The bulb flickers 
or the brightness is 
not stable. 

The wire doesn’t connect all right. 

Connect correctly. 

 

Содержание MRJ-01

Страница 1: ...1 PROFESSIONAL MICROSCOPE Model MRJ 01 Monocular MRJ 03 Binocular MRJ 03D Dual View MRJ 03T Trinocular ...

Страница 2: ...ents MRJ 1 Before use 3 2 Diagram 4 3 Assemble 6 4 Operation 6 5 General specification 10 6 Configuration drawing 11 7 Specification 12 8 Troubleshooting 13 9 Trinocular Dual View microscope 15 10 Warranty 15 ...

Страница 3: ... Lamp selected only 6V 20W Halogen Bulb 5 This microscope has a built in dual voltage transformer and can be used with wide voltage range from 90 to 240V 6 Use only electrical plug supplied with the microscope 1 2 Maintenance 1 Wipe the lens gently with a soft lens tissue After every use with oil or fingerprints clean the lens surfaces with lens paper or lens paper moistened with a little rubbing ...

Страница 4: ...MRJ MRJ Intermediate Biological Microscope Diagram Eyepieces Head lock screw Arm Head Nosepiece Objectives Stage Stop Stage Course Focus Condenser Light Variable light control knob Fine Focus Condenser up down knob ...

Страница 5: ...5 2 Diagram MRJ MRJ Biological Microscope Diagram Course Focus Fine Focus Mechanical Stage X Axis knob Condenser Y Axis knob Condenser up down knob Condenser center adjustment screw Aperture Diaphragm ...

Страница 6: ...e plug into an outlet and turn on 2 Adjust the light adjustment knob until you get the suitable brightness See fig 1 Fig 1 4 2 Set the specimen slide 1 Push the arm of the specimen holder back 2 Insert the slide with the cover glass facing up and allow the arm of the specimen holder forward See fig 2 3 Rotate the X and Y axis knobs on the mechanical stage to move the specimen to the center the cen...

Страница 7: ...llary distance When observing with both eyes hold the bases of the base of the eyepieces and rotate them around the axis until there is only one field of view See fig 4 On the left and right eyepiece base should point to the same number The number is the interpupillary distance Fig 4 4 5 Adjust diopters Rotate one eyepiece to the 0 position using the white thread This should be done to your domina...

Страница 8: ...e See fig 7 When the shadow around the field of view of the eyepiece is symmetrical it shows the condenser has been centered correctly See fig 7 6 Now you can enlarge the aperture diaphragm and remove the shadow Fig 7 Outside edge of objective 4 7 Adjust aperture diaphragm The aperture diaphragm decides the numerical aperture of the illumination If the N A of illumination is matching with N A of t...

Страница 9: ...oil does not come off easily 4 9 Use the filter Filter can make the background softer fig 11 Filter holder There is a filters blue You also place the Polarizer in this position when using the optional simple polarizing set Fig 11 Fig 12 4 10 Replace the halogen bulb 1 Turn off the power supply before replacing the bulb 2 Loosen the screw of the lamp holder base and open the door to the lamp holder...

Страница 10: ...achromatic objective 40X Objective Plan achromatic objective 100X 45 Monocular Head 30 ocular head Bin 30 lar head Trinocu Head 30 Dual Binocular head Nosepiece Rev ple nosepiece ersed quadru Stand C Stage 140 x 135mm mechanical stage N A 1 25 abbe condenser N A 1 25 five position phase Si ase contrast ngle special ph Single special phase contrast Single special phase contrast Power 90 ide voltage...

Страница 11: ...ular Body Center Eyepiece 15X Eyepiece Filter Polarizer Binocular 10X Eyepiece N A 1 25 Oil dark field condenser Blue Green Mechanical moving stage for left N A 0 9 Dry dark field condenser N A 1 25 Abbe achromatic condenser Phase contrast objective Achromatic objective N A 1 25 Phase contrast condenser ...

Страница 12: ... adjustment left and right hand for option Condenser N A 1 25 achromatic condenser N A 1 25 five position phase contrast condenser not included N A 0 9 dry dark field condenser not included N A 1 25 oil dark field condenser not included Illumination 90 240V wide voltage Built in Koehler illuminator system 6V 20W halogen bulb Pre centered a continuous adjustment of brightness 7 2 Parameter of objec...

Страница 13: ...ss 0 17mm The cover glass faces down Put the cover glass to face up The immersion oil has accumulated on the dry objective Clean thoroughly The immersion oil is not used for oil objective 100XR Use immersion oil Air bubble in the immersion Get rid of the air bubble Use wrong immersion oil Use correct one The aperture is not opened correctly Adjust the iris diaphragm Stain or dust has accumulated o...

Страница 14: ...arse focusing knob can t decline The base of the condenser is too low Raise the base The slide is not fixed correctly Adjust it correctly 7 Can not move the slide smoothly The movable specimen holder is not fixed properly Adjust it correctly 8 The image moves obviously when touching the stage The stage is fastened incorrectly Fasten the stage correctly Electrical No power supply Check the connecti...

Страница 15: ... year for electrical components It does not cover damage resulting from abuse or misuse repairs or alterations performed by other than authorized repair technicians or damage occurring in transit If you have questions concerning this product or warranty contact the dealer from whom it was purchased For warranty service microscope should be well packed to avoid damage in transit preferably in origi...

Отзывы: