9 Technical Specifications
_____________________________________________________________________________________________________________________________________
9-27
Measurement of EMI Immunity
Conforming EN 50082-1,
European Standard
,
Electromagnetic Compatibility - Generic Immunity Standard
Generic Standard Class: Residential, Commercial, or Light Industry.
Additional 5017 specifications under EMC test conditions
ENV 50140
EMC Immunity against radiated EMI (Electromagnetic Interference).
EMC Basic Directive, 30-1000MHz. Residential environment
EMI Radiation: 3V/m, 80%AM - f.e. Handy in a distance of 3-5m
5017 Specifications
min. +/- 0.1 % F. S. (Full Scale)
ENV 50141
EMC Immunity against conducted EMI
EMC Basic Directive, 150 kHz - 80 MHz
EMI Source: 3V
RMS
on open wires.
5017 Specifications
min. +/- 0.1% F.S.
EN 61000-4-2
ESD Immunity Part 4-2
ESD: Air or Contact Dis/-8kV / 4 kV - Protection Class 2.
5017 Specifications
min. +/- 0.1% F.S.
EN 61000-4-4
Burst and electrical fast transient immunity Part 4-4
Environment: Industrial Area 2kVpk - Protection Class 3.
5017 Specifications
min. +/- 0.1% F.S.
EN 61000-4-5
Surge Immunity Test, Part 4-5
Environment: Industrial Area 2kV non symm. - Protection Class 3.
5017 Specifications
min. +/- 0.1% F.S.