35
6-2 Pressure parameter calibration
Pressure parameter calibration should calibrate zero first then span, this sequence can not be reversed.
(1) Multimeter connects with J6 pin S and G on main control board, refer to fig 6
‐
5, adjust side Z until the
output voltage is 10 mV . This operation process is zero calibration
Fig
6-5 Wiring layout for pressure calibration
(2) Connect J6 pin G and J with a mini jumper, adjust side S until the output voltage is 4.00 V then entering
setting interface and press button RP, refer to fig 6
‐
4, finally click button OK to save calibration value. If the
value displayed on screen is not right, pressure parameter should be calibrated again.
7 Wiring diagram