Fault

Probable cause

Solution

Turntable does not work

Photoelectric sensor fault

Motor fault

Main control board error

Replace photoelectric sensor

Replace motor

Replace main control board

Touch screen fault

Resistance broken

No calibration

Main control board error

Replace touch screen

Calibrate touch function

Replace main control board

Garble code on screen

Bad data lines connection

LCD fault

Main control board error

Re

installation the data lines

Replace LCD

Replace main control board

No video show

No video signal

No working voltage

Monitor screen fault

Replace pin hole camera

Check 12V power and its wiring

Replace monitor screen

No response for ‘auto open
door’ function

Exhaust fan is open

Main control board error

No 24V work voltage

Electromagnet fault

Axis of electromagnet decentration

Stop exhaust fan

Replace main control board

Replace power adapter (24V)

Replace electromagnet

Re

mount the door mechanism

Warning :

“Time

out”

No microwave

Wrong microwave power setting

Poor microwave absorb ability

Heating time is short

Check microwave generated components

Set right microwave power

Increase microwave power

Prolong heating time and set multi

step

Warning:

“Temperature error”

Air leakage

Pressure sensor fault

Main control board error

Power is setted too large

Check the air impermeability

Replace pressure sensor

Replace main control board

Set related microwave power

Warning:

‘Pressure limited’

Power is setted too large

Pressure setting value is lower

Samples have good microwave
absorbing ability

Pressure sensor fault

Reduce microwave power

Increase setting value

Reduce microwave power

Replace pressure sensor

Warning:

‘Pressure control’

Lower pressure setting value

Main control board error

Increase setting value

Replace main control board

Warning:

‘Turntable fault’

Photoelectric sensor fault

Main control board error

Turntable motor fault

Replace Photoelectric sensor

Replace main control board

Replace motor

30

TOPEX service manual

4-1 HV components measuring method

Measuring the HV components input or output terminals with the power on is forbidden.These HV

components can be estimated by measuring its resistance.Test behavior can only be operated 5 minutes

latter after power off.

4-1-1 Filament transformer

Disconnect the primary and secondary wiring terminals,the resistance of primary is about 20

Ω

and the

secondary is 1

Ω

normally. If resistance is right, we can eliminate the doubt of filament transformer fault

4-1-2 HV transformer

Disconnect the primary and secondary wiring terminals,the resistance of primary is about 1.5

Ω

and the

secondary is nearly 100

Ω

normally (between secondary and grounding).

Fault type: open circuit

4-1-3 HV capacitor

The resistance between two terminals is about 10M

Ω

.

Lower or higher resistance indicate the capacitor has been break down.

Resistance should escalate while the measuring process, if the resistance is directly fixed at 10M

Ω

, no

escalation process, that indicates the HV capacitor has been break down.

Fault type: Break down

4-1-4 HV diode

Take down HV diode

If no microwave generated and the surface temperature is too high that indicates HV diode has been

break down.

The resistance between anode and cathode is about 30 to 70M

Ω

.

The resistance between cathode and anode is infinite.

If the forward direction resistance is too high or reverse is too low beyond the standard indicates break

down.

The anode must be connected with HV capacitor and reliable grounding for cathode.

4-1-5 Magnetron

Resistance of filament cathode is nearly 1

Ω

.

Resistance between filament cathode and grounding is infinite.

Magnet flaws on magnetron is not allowed.

Содержание TOPEX

Страница 1: ...cess of maintenance Before troubleshooting maintenance engineer should be familiar with instrument wiring diagram fully understand the components work voltage and working mechanism Please give special...

Страница 2: ...0 3 8 Turntable motor replacement 21 3 9 LCD replacement 23 3 10 Built in camera and monitor replacement 24 3 11 LED backlight replacement 26 3 12 Push style electromagnet replacement 26 4 Trouble sho...

Страница 3: ...ure Microwave can only be generated when the real time temperature is not beyond setting value 1 4 Safety standards Troubleshooting should not be dealed with power on please refer to the following rul...

Страница 4: ...er to fig 2 2 Fig 2 2 Fuse holders We define the above thyristor belongs to fisrt group and the bellow one belongs to the second group Other components like magnetrons high voltage items can be define...

Страница 5: ...3 4 LED light 00cz021 5 Pressure plate 26aa015 6 PCB board Built in monitor 7 monitor 00fa704 8 Power PCB Built in monitor 9 Holder 26aa014 10 TOPEX service manual 2 3 Bottom view Fig 2 3 bottom compo...

Страница 6: ...11 2 4 Main control board layout No Describe Inventory code 5 Photoelectric sensor 26tj205 6 bracket 23aa020 7 bracket 09aa042 12 TOPEX service manual...

Страница 7: ...ce of the magnetron Note Exhaust duct is mounted with magnetron please don t take down when replace the magnetron Fig 3 2 magnetron The second magnetron The first magnetron Nuts Exhaust duct Temperatu...

Страница 8: ...like fig 3 8 take down four fixed nuts and disconnect the cord We can do replacement process Two HV transformers are installed as the same way Wiring layout We normally do not stipulate the input wiri...

Страница 9: ...tor its conductive feature is controlled by signal Control signal is afford from JP4 15 20 pins pin 15 17 and 19 belong to the first thyristor pin 16 18 and 20 belong to the second one the range of co...

Страница 10: ...ng silicone 4 Remount the exhaust fan fixed by three M5 nuts refer to fig 3 16 5 Restore wiring connection Fig 3 17 Sealing silicone Fig 3 18 Connection bracket Exhaust fan wiring diagram 20 TOPEX ser...

Страница 11: ...nuts Turntable motor is a subassembly gathered with metal bracket photoelectric sensor and grating Note 1 Instrument should be overturned take down the black cover 2 Take the turntable and vessels out...

Страница 12: ...ed by four M3 6 screws data communication type is serial communication via a LCD connector connected with main control board refer to wiring diagram There are four points need to pay more attention wh...

Страница 13: ...ionship is described like fig 3 30 Fig 3 30 Monitor screen subassembly Note Monitor and its PCB connection style is push pull way strange form LCD refer to fig 3 31 Fig 3 31 Push pull style 26 TOPEX s...

Страница 14: ...fan does not work Replace related temperature Temperature value unstable Microwave interference Temperature socket loosen Temperature sensor Main control board error Check door gap Re mount the temper...

Страница 15: ...board error Increase setting value Replace main control board Warning Turntable fault Photoelectric sensor fault Main control board error Turntable motor fault Replace Photoelectric sensor Replace mai...

Страница 16: ...rowave power match Thyristor can not control its output voltage may cause this fault there are three cause reasons 1 Power control signal afford by main control board is error 2 Thyristor is broken 3...

Страница 17: ...at is gray status represents lowercase and highlight represents uppercase refer to fig 6 2 and 6 3 Fig 6 2 Lowercase status Status button 34 TOPEX service manual Fig 6 3 Uppercase status Fig 6 4 Tempe...

Страница 18: ...l the output voltage is 10 mV This operation process is zero calibration Fig 6 5 Wiring layout for pressure calibration 2 Connect J6 pin G and J with a mini jumper adjust side S until the output volta...

Страница 19: ...Preekem Scientific Instruments Co Ltd No 2 Building No 4299 Jindu Road Shanghai 201108 China Tel 86 021 5442 7057 Fax 86 021 5442 7063 E mail Support preekem com Service manual Ver 1 0...

Отзывы: