BLENDEr

HELIUM-OxygEn

COMPONENT dESCRIPTION

ITEM

 

DESCRIPTION

A

Oxygen Concentration dial

 A dial used for selecting oxygen concentrations between 20%

 

-100% or 30%

 

-100%. The F

I

O

2

 scale is used for reference 

only. 

The actual 

F

I

O

2

 must be verified with an Alarmed 

Oxygen Monitor / Analyzer.

This dial does not rotate 360°.

 The dial starts at 20% or 

30% and ends at 100%.

 

B

Primary Outlet Port 

A male DISS oxygen fitting with check valve that delivers flow 

when engaged to any controlling device, such as a flowmeter.

 

C

Auxiliary Outlet Port 

 A male DISS oxygen fitting with check valve that delivers flow 

when engaged to any controlling device, such as a flowmeter. 

This outlet is equipped with a bleed valve that allows the user 

to control if the bleed is ON or OFF. With the bleed in the “ON” 

position, this outlet delivers accurate oxygen concentrations 

in the following flows: 

Model 

Flow Range

High Flow 

2 - 100 lpm

Low Flow 

0 -

  

30 lpm

d

Auxiliary Bleed Collar

 

 The collar is used to engage and disengage the bleed. The 

bleed is necessary to maintain accurate F

I

O

2

 Concentration 

below 15 lpm for the High Flow and 

 3 lpm for the Low Flow. 

To activate the bleed, slide and rotate the knurled collar back 

until it contacts the cover. To deactivate the bleed, pull and 

rotate collar away from cover until 

bleed flow valve is closed.

E

Oxygen Inlet Fitting

 A female DISS or NIST oxygen fitting with one way valve that 

is used to connect an oxygen supply hose.

F

Heliox Inlet Fitting

 

A male DISS or NIST heliox fitting with one way valve that 

is used to connect a heliox supply hose.

G

Alarm 

An audible alarm that sounds due to an excessive 

pressure drop or deletion of either gas supply. 

H

Rear Slide Mount 

with dove tail.

Содержание HELIO2 PM5500 Series

Страница 1: ...ctions Model No PM5400 Series shown PM5500 Series SAVE THESE INSTRUCTIONS CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 No...

Страница 2: ...ixtures of 20 oxygen 80 helium and 30 to 100 for heliox tank mixtures of 30 oxygen 70 helium The Helium and Oxygen Blender is a restricted medical device intended for use by qualified and trained pers...

Страница 3: ...hread psi Pounds Per Square Inch lpm Liters Per Minute SAFETY Information Warnings and Cautions DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious...

Страница 4: ...ender when alarm is sounding DO NOT use oil in or around the Blender DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender DO NOT use near any type of flame or flammable exp...

Страница 5: ...Flow Loss of Helium or Oxygen supply 85 lpm 45 lpm Bypass Alarm Activation 50 psi 3 45 bar 60 psi 4 14 bar 50 psi 3 45 bar 60 psi 4 14 bar 13 25 psi 16 24 psi 18 22 psi 16 24 psi 0 9 1 7 bar 1 1 1 65...

Страница 6: ...Helium Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195 The alarmed Oxygen Monitor Analyzer should comply with ISO 7767 to meet CE requirement Dryness and Composition for inlet gase...

Страница 7: ...6 HELIO2 Blender HELIUM Oxygen Diagrams CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc PM5400 and PM5500 Models A B F E D C H G...

Страница 8: ...or OFF With the bleed in the ON position this outlet delivers accurate oxygen concentrations in the following flows Model Flow Range High Flow 2 100 lpm Low Flow 0 30 lpm D Auxiliary Bleed Collar The...

Страница 9: ...e auxiliary flow outlet maintains the same flow capacity and FIO2 accuracy as the Primary Outlets with Bleed Valve not engaged When the bleed flow is activated some of the heliox mixture will vent to...

Страница 10: ...damaged 1 Secure Blender to wall or pole mount bracket 2 Connect heliox and oxygen supply lines from HELIO2 Blender to heliox cylinder and oxygen supply 3 Connect oxygen flowmeter to Blender outlet 4...

Страница 11: ...Oxide DO NOT clean with aromatic hydrocarbons 1 Disconnect all gas connections and equipment before cleaning 2 Clean exterior surfaces with a cloth dampened with mild detergent and water 3 Wipe dry wi...

Страница 12: ...0 18 32 4 28 8 25 2 23 4 21 6 21 2 20 7 18 4 18 0 19 34 2 30 4 26 6 24 7 22 8 22 4 21 9 19 4 19 0 20 36 0 32 0 28 0 26 0 24 0 23 6 23 0 20 4 20 0 21 37 8 33 6 29 4 27 3 25 2 24 8 24 2 21 4 21 0 22 39...

Страница 13: ...rned Goods Authorization RGA number contact Precision Medical Inc All returns must be packaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in...

Страница 14: ...inated or heliox concentration incorrect 5 Downstream device causing back flow or restricted flow 6 Supply pressure imbalanced 1 Use auxiliary outlet engage bleed 2 Recalibrate alarmed Oxygen Monitor...

Страница 15: ...ke oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and...

Страница 16: ...odel s PM5470EN PM5480EN PM5570EN PM5580EN MDD Class IIb Classification criteria Clause 3 2 Rule 11 of Annex IX of MDD As delivered the object of the declaration described above is in conformity with...

Отзывы: