Precision Medical HELIO2 PM5500 Series Скачать руководство пользователя страница 14

13

H

ELI

O

2

BLENDEr

HELIUM-OxygEn

TROUBLESHOOTING

If the

 

H

ELI

O

2

 

Blender

 fails to function, consult the Troubleshooting 

Guide below.

If problem cannot be solved by using Troubleshooting Guide, refer to the

 

H

ELI

O

2

 

Blender

 Service Manual (P/N 506124) available on the Internet; 

www.precisionmedical.com or consult your Provider.

Problem

Probable Cause

Remedy

Oxygen 

concentration 

discrepancy 

between Blender 

setting and 

alarmed Oxygen 

Monitor / Analyzer 

(greater than 3%)

1. •

HIgH

 flow model, 

flow requirement 

below 15 lpm.

LOW

 flow model, 

flow requirement 

below 3 lpm.

2. Alarmed Oxygen 

Monitor / Analyzer 

inaccurate

3. Low flow bleed 

obstructed

4. Gas supply 

contaminated or 

heliox concentration 

incorrect

5. Downstream device 

causing back flow 

or restricted flow

6. Supply pressure 

imbalanced

1. Use auxiliary outlet & 

engage bleed

2. Recalibrate alarmed 

Oxygen Monitor / 

Analyzer or Verify 

with second alarmed 

Oxygen Monitor / 

Analyzer

3. Remove obstruction

4. Check gas sources 

with calibrated alarmed 

Oxygen Monitor / 

Analyzer to

 confirm 

oxygen is 100% and 

verify heliox tank content

5. Isolate Blender. Check 

oxygen concentration at 

Blender Outlets

6. Assure heliox and 

oxygen inlets pressures 

are within 10 psi

No flow at

Blender outlets

1. Gas sources turned 

“OFF”

2. Gas sources not 

connected

1. Turn gas sources “ON”

2. Connect gas sources

Alarm sounding

1. Difference between 

oxygen and heliox 

inlet pressures 

greater than specified

1. Correct pressure 

difference until heliox 

and oxygen pressures 

are within specification

Содержание HELIO2 PM5500 Series

Страница 1: ...ctions Model No PM5400 Series shown PM5500 Series SAVE THESE INSTRUCTIONS CAUTION Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 No...

Страница 2: ...ixtures of 20 oxygen 80 helium and 30 to 100 for heliox tank mixtures of 30 oxygen 70 helium The Helium and Oxygen Blender is a restricted medical device intended for use by qualified and trained pers...

Страница 3: ...hread psi Pounds Per Square Inch lpm Liters Per Minute SAFETY Information Warnings and Cautions DANGER Indicates an imminently hazardous situation which if not avoided will result in death or serious...

Страница 4: ...ender when alarm is sounding DO NOT use oil in or around the Blender DO NOT occlude or obstruct the bleed port on the auxiliary outlet of the Blender DO NOT use near any type of flame or flammable exp...

Страница 5: ...Flow Loss of Helium or Oxygen supply 85 lpm 45 lpm Bypass Alarm Activation 50 psi 3 45 bar 60 psi 4 14 bar 50 psi 3 45 bar 60 psi 4 14 bar 13 25 psi 16 24 psi 18 22 psi 16 24 psi 0 9 1 7 bar 1 1 1 65...

Страница 6: ...Helium Oxygen Blender reverse gas flow complies with clause 6 of ISO 11195 The alarmed Oxygen Monitor Analyzer should comply with ISO 7767 to meet CE requirement Dryness and Composition for inlet gase...

Страница 7: ...6 HELIO2 Blender HELIUM Oxygen Diagrams CAUTION Missing or illegible labels must be replaced contact Precision Medical Inc PM5400 and PM5500 Models A B F E D C H G...

Страница 8: ...or OFF With the bleed in the ON position this outlet delivers accurate oxygen concentrations in the following flows Model Flow Range High Flow 2 100 lpm Low Flow 0 30 lpm D Auxiliary Bleed Collar The...

Страница 9: ...e auxiliary flow outlet maintains the same flow capacity and FIO2 accuracy as the Primary Outlets with Bleed Valve not engaged When the bleed flow is activated some of the heliox mixture will vent to...

Страница 10: ...damaged 1 Secure Blender to wall or pole mount bracket 2 Connect heliox and oxygen supply lines from HELIO2 Blender to heliox cylinder and oxygen supply 3 Connect oxygen flowmeter to Blender outlet 4...

Страница 11: ...Oxide DO NOT clean with aromatic hydrocarbons 1 Disconnect all gas connections and equipment before cleaning 2 Clean exterior surfaces with a cloth dampened with mild detergent and water 3 Wipe dry wi...

Страница 12: ...0 18 32 4 28 8 25 2 23 4 21 6 21 2 20 7 18 4 18 0 19 34 2 30 4 26 6 24 7 22 8 22 4 21 9 19 4 19 0 20 36 0 32 0 28 0 26 0 24 0 23 6 23 0 20 4 20 0 21 37 8 33 6 29 4 27 3 25 2 24 8 24 2 21 4 21 0 22 39...

Страница 13: ...rned Goods Authorization RGA number contact Precision Medical Inc All returns must be packaged in sealed containers to prevent damage Precision Medical Inc will not be responsible for goods damaged in...

Страница 14: ...inated or heliox concentration incorrect 5 Downstream device causing back flow or restricted flow 6 Supply pressure imbalanced 1 Use auxiliary outlet engage bleed 2 Recalibrate alarmed Oxygen Monitor...

Страница 15: ...ke oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale Thus this writing is a final complete and...

Страница 16: ...odel s PM5470EN PM5480EN PM5570EN PM5580EN MDD Class IIb Classification criteria Clause 3 2 Rule 11 of Annex IX of MDD As delivered the object of the declaration described above is in conformity with...

Отзывы: